SwePub
Sök i LIBRIS databas

  Extended search

id:"swepub:oai:research.chalmers.se:cb30f97e-d171-4072-9510-06d9c366eb18"
 

Search: id:"swepub:oai:research.chalmers.se:cb30f97e-d171-4072-9510-06d9c366eb18" > Dynamic Range Expan...

  • 1 of 1
  • Previous record
  • Next record
  •    To hitlist

Dynamic Range Expansion of the C-Reactive Protein Quantification with a Tandem Giant Magnetoresistance Biosensor

Meng, Fanda, 1987 (author)
Shandong Qianfoshan Hospital,Chalmers tekniska högskola,Chalmers University of Technology
Zhang, Lei (author)
Chinese Academy of Sciences,Dongguan Bosh Biotechnologies, Ltd.
Huo, Weisong (author)
Dongguan Bosh Biotechnologies, Ltd.
show more...
Lian, Jie (author)
People's Public Security University of China
Jesorka, Aldo, 1967 (author)
Chalmers tekniska högskola,Chalmers University of Technology
Shi, Xizeng (author)
Dongguan Bosh Biotechnologies, Ltd.
Gao, Yunhua (author)
Chinese Academy of Sciences
show less...
 (creator_code:org_t)
2021-05-07
2021
English.
In: ACS Omega. - : American Chemical Society (ACS). - 2470-1343. ; 6:19, s. 12923-12930
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • In this study, we report a convenient analytical method for a full-range quantification of the C-reactive protein (CRP), a blood biomarker of infection and cardiovascular events. We determine CRP over the entire diagnostically relevant concentration range in undiluted human blood serum in a single test, using a tandem giant magnetoresistance (GMR) sensor. The tandem principle combines a sandwich assay and a competitive assay, which allows for the discrimination of the concentration values resulting from the multivalued dose-response curve ("Hook"effect), which characterizes the one-step sandwich assay at high CRP concentrations. The sensor covers a linear detection range for CRP concentration from 3 ng/mL to 350 μg/mL, the detection limit (s/n = 3) is 1 ng/mL. The prominent features of the chip-based method are its expanded dynamic range and low sample volume (50 μL), and the need for a short measurement time of 15 min. These figures of merit, in addition to the low detection limit equal to the established assay instrumentation, make it a viable candidate for use in point-of-care diagnostics.

Subject headings

NATURVETENSKAP  -- Kemi -- Analytisk kemi (hsv//swe)
NATURAL SCIENCES  -- Chemical Sciences -- Analytical Chemistry (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Klinisk laboratoriemedicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Clinical Laboratory Medicine (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)

Publication and Content Type

art (subject category)
ref (subject category)

Find in a library

  • ACS Omega (Search for host publication in LIBRIS)

To the university's database