Comparison of three methods for determining CT dose profile: presenting the tritium method

Hansson, Jonny (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
Eriksson, Simone (författare)
Thilander-Klang, Anne, 1959 (författare)
Gothenburg University,Göteborgs universitet,Extern,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,External,Institute of Clinical Sciences, Department of Radiation Physics
visa fler...
Båth, Magnus, 1974 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
visa färre...
 (creator_code:org_t)
2010-03-03
2010
Engelska.
Ingår i: Radiation protection dosimetry. - : Oxford University Press (OUP). - 1742-3406 .- 0144-8420. ; 139:1-3, s. 434-438
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The purpose of the present work was to describe a method of using an imaging plate from a computed radiography system to determine the computed tomography (CT) dose profile (the tritium method) and to compare this method with point-dose measurements using a solid-state detector (CT Dose Profiler; RTI Electronics, Mölndal, Sweden) and the indirect method of comparing the air kerma-length product (P(KL)) at different beam collimations. The three methods were used to determine the full width at half maximum (FWHM) of the dose profile of a multi-slice CT at different nominal beam collimations. For all beam collimations, the obtained deviation between the tritium method and the CT Dose Profiler was smaller than 0.1 mm. The maximum relative error was 2 %. For the P(KL) method, the deviation from the CT Dose Profiler was between 0.2 and 0.4 mm, resulting in a relative error larger than 10 % for the smallest beam collimation even after normalisation to a known FWHM. In conclusion, the proposed method of using an imaging plate to determine the FWHM of the CT dose profile has a high accuracy and shows good agreement with the more advanced method of point-dose measurements using a solid-state detector.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy