Quantifying the Effect of 3T Magnetic Resonance Imaging Residual System Distortions and Patient-Induced Susceptibility Distortions on Radiation Therapy Treatment Planning for Prostate Cancer

Adjeiwaah, Mary (författare)
Umeå universitet,Radiofysik,Umea Univ, Dept Radiat Sci, SE-90187 Umea, Sweden.
Bylund, Mikael (författare)
Umeå universitet,Radiofysik,Umea Univ, Dept Radiat Sci, SE-90187 Umea, Sweden.
Lundman, Josef A. (författare)
Umeå universitet,Radiofysik,Umea Univ, Dept Radiat Sci, SE-90187 Umea, Sweden.
visa fler...
Thellenberg Karlsson, Camilla (författare)
Umeå universitet,Onkologi,Umea Univ, Dept Radiat Sci, SE-90187 Umea, Sweden.
Jonsson, Joakim H. (författare)
Umeå universitet,Radiofysik,Umea Univ, Dept Radiat Sci, SE-90187 Umea, Sweden.
Nyholm, Tufve (författare)
Uppsala universitet,Umeå universitet,Radiofysik,Medical Radiation Physics, Department of Immunology, Genetics and Pathology, Uppsala University, Uppsala, Sweden,Medicinsk strålningsvetenskap,Umea Univ, Dept Radiat Sci, SE-90187 Umea, Sweden
visa färre...
 (creator_code:org_t)
Elsevier, 2018
2018
Engelska.
Ingår i: International Journal of Radiation Oncology, Biology, Physics. - : Elsevier. - 0360-3016 .- 1879-355X. ; 100:2, s. 317-324
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose: To investigate the effect of magnetic resonance system- and patient-induced susceptibility distortions from a 3T scanner on dose distributions for prostate cancers.Methods and Materials: Combined displacement fields from the residual system and patient-induced susceptibility distortions were used to distort 17 prostate patient CT images. VMAT dose plans were initially optimized on distorted CT images and the plan parameters transferred to the original patient CT images to calculate a new dose distribution.Results: Maximum residual mean distortions of 3.19 mm at a radial distance of 25 cm and maximum mean patient-induced susceptibility shifts of 5.8 mm were found using the lowest bandwidth of 122 Hz per pixel. There was a dose difference of <0.5% between distorted and undistorted treatment plans. The 90% confidence intervals of the mean difference between the dCT and CT treatment plans were all within an equivalence interval of (−0.5, 0.5) for all investigated plan quality measures.Conclusions: Patient-induced susceptibility distortions at high field strengths in closed bore magnetic resonance scanners are larger than residual system distortions after using vendor-supplied 3-dimensional correction for the delineated regions studied. However, errors in dose due to disturbed patient outline and shifts caused by patient-induced susceptibility effects are below 0.5%.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy