Evaluating the accuracy and precision of a two-compartment Kärger model using Monte Carlo simulations.

Nilsson, Markus (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Alerstam, Erik (författare)
Wirestam, Ronnie (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa fler...
Ståhlberg, Freddy (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk strålningsfysik, Lund,Institutionen för kliniska vetenskaper, Lund,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Medical Radiation Physics, Lund,Department of Clinical Sciences, Lund
Brockstedt, Sara (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Lätt, Jimmy (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
visa färre...
 (creator_code:org_t)
Elsevier BV, 2010
2010
Engelska.
Ingår i: Journal of Magnetic Resonance. - : Elsevier BV. - 1096-0856 .- 1090-7807. ; 206:1, s. 59-67
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Specific parameters of the neuronal tissue microstructure, such as axonal diameters, membrane permeability and intracellular water fractions are assessable using diffusion MRI. These parameters are commonly estimated using analytical models, which may introduce bias in the estimated parameters due to the approximations made when deriving the models. As an alternative to using analytical models, a database of signal curves generated by fast Monte Carlo simulations can be employed. Simulated diffusion MRI measurements were generated and evaluated using the two-compartment Kärger model as well as the simulation model based on a database containing signal curves from approximately 60000 simulations performed with different combinations of microstructural parameters. A protocol based on a pulsed gradient spin echo sequence with diffusion times of 30 and 60ms and with gradient amplitudes obtainable with a clinical MRI scanner was employed for the investigations. When using the analytical model, a major negative bias (up to approximately 25%) in the estimated intracellular volume fraction was observed for short exchange times, while almost no bias was seen for the simulation model. In general, the simulation model improved the accuracy of the estimated parameters as compared to the analytical model, except for the exchange time parameter.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy