Search: WFRF:(Jonson Lars) > English > A new CT method for...

A new CT method for measuring cup orientation after total hip arthroplasty : A study of 10 patients

Olivecrona, Henrik (author)
Karolinska Institute, Department of Molecular Medicine and Surgery
Weidenhielm, Lars (author)
Karolinska Institutet
Olivecrona, Lotta (author)
Karolinska Institutet, Department of Clinical Science, Intervention and Technology, Division of Radiology
show more...
Beckman, M. O. (author)
Karolinska Institutet
Stark, Andreas (author)
Karolinska Institutet
Noz, Marilyn E. (author)
New York University, Department of Radiology,Nuclear Medicine
Maguire Jr., Gerald Q. (author)
KTH,Mikroelektronik och informationsteknik, IMIT
Zeleznik, Michael P. (author)
RAHD Oncology Products, St. Loius MO, USA
Svensson, Lars E. (author)
KTH,Matematik
Jonson, T. (author)
show less...
 (creator_code:org_t)
2009-07-08
2004
English.
In: Acta Orthopaedica Scandinavica. - : Medical Journals Sweden AB. - 0001-6470. ; 75:3, s. 252-260
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • Background It is difficult to assess the orientation of the acetabular component on routine radiographs. We present a method for determining the spatial orientation of the acetabular component after total hip arthroplasty (THA) using computed tomography. Patients and methods Two CT-scans, 10 min apart, were obtained from each of 10 patients after THA. Using locally developed software, two independent examiners measured the orientation of the acetabular component in relation to the pelvis. The measurements were repeated after one week. To be independent of the patient position during scanning, the method involved two steps. Firstly, a 3D volumetric image of the pelvis was brought into a standard pelvic orientation, then the orientation of the acetabular component was measured. The orientation of the acetabular component was expressed as operative anteversion and inclination relative to an internal pelvic reference coordinate system. To evaluate precision, we compared measurements across pairs of CT volumes between observers and trials. Results Mean absolute interobserver angle error was 2.3degrees for anteversion (range 0-6.6degrees), and 1.1degrees for inclination (range 0-4.6degrees). For interobserver measurements, the precision, defined as one standard deviation, was 2.9degrees for anteversion, and 1.5degrees for inclination. A Student's West showed that the overall differences between the examiners, trials, and cases were not significant. Data were normally distributed and were not dependent on examiner or trial. Interpretation We conclude that the implant angles of the acetabular component in relation to the pelvis could be detected repeatedly using CT, independently of patient positioning.

Subject headings

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Keyword

acetabular component migration
semiautomated program
fusion
repeatability
dislocation
replacement
prosthesis

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view