Sökning: onr:"swepub:oai:DiVA.org:kth-188117" > Superficial vessel ...

Superficial vessel reconstruction with a multiview camera system

Maria Marreiros, Filipe Miguel, 1978- (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten
Rossitti, Sandro (författare)
Linköpings universitet,Östergötlands Läns Landsting,Neurokirurgiska kliniken US,Avdelningen för neuro- och inflammationsvetenskap,Medicinska fakulteten
Karlsson, Per (författare)
Östergötlands Läns Landsting,Neurokirurgiska kliniken US
visa fler...
Wang, Chunliang (författare)
Linköpings universitet,KTH,Medicinsk bildbehandling och visualisering,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Royal Institute of Technology, School of Technology and Health, Alfred Nobels Allé 10, Huddinge
Gustafsson, Torbjorn (författare)
XM Reality AB, Linköping, Sweden
Carleberg, Per (författare)
XM Reality AB, Linköping, Sweden
Smedby, Örjan (författare)
Linköpings universitet,KTH,Medicinsk bildbehandling och visualisering,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Medie- och Informationsteknik,Tekniska fakulteten,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Region Östergötland, Röntgenkliniken i Linköping,Royal Institute of Technology, School of Technology and Health, Alfred Nobels Allé 10, Huddinge
visa färre...
 (creator_code:org_t)
SPIE - International Society for Optical Engineering, 2016
2016
Engelska.
Ingår i: Journal of Medical Imaging. - : SPIE - International Society for Optical Engineering. - 2329-4302 .- 2329-4310. ; 3:1
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • We aim at reconstructing superficial vessels of the brain. Ultimately, they will serve to guide the deformation methods to compensate for the brain shift. A pipeline for three-dimensional (3-D) vessel reconstruction using three mono-complementary metal-oxide semiconductor cameras has been developed. Vessel centerlines are manually selected in the images. Using the properties of the Hessian matrix, the centerline points are assigned direction information. For correspondence matching, a combination of methods was used. The process starts with epipolar and spatial coherence constraints (geometrical constraints), followed by relaxation labeling and an iterative filtering where the 3-D points are compared to surfaces obtained using the thin-plate spline with decreasing relaxation parameter. Finally, the points are shifted to their local centroid position. Evaluation in virtual, phantom, and experimental images, including intraoperative data from patient experiments, shows that, with appropriate camera positions, the error estimates (root-mean square error and mean error) are similar to 1 mm.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)
NATURVETENSKAP  -- Data- och informationsvetenskap -- Datorseende och robotik (hsv//swe)
NATURAL SCIENCES  -- Computer and Information Sciences -- Computer Vision and Robotics (hsv//eng)

Nyckelord

brain shift correction
three-dimensional vessel tracking
multiview geometry

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy