Blood plasma coagulation studied by surface plasmon resonance

Vikinge, Trine P. (author)
Linköpings universitet,Tillämpad Fysik,Tekniska högskolan
Hansson, Kenny, 1972- (author)
Linköpings universitet,Klinisk kemi,Hälsouniversitetet
Benesch, Johan (author)
Linköpings universitet,Tillämpad Fysik,Tekniska högskolan
show more...
Johansen, Knut (author)
Linköpings universitet,Tillämpad Fysik,Tekniska högskolan
Rånby, Mats (author)
Linköpings universitet,Klinisk kemi,Hälsouniversitetet
Lindahl, Tomas, 1954- (author)
Linköpings universitet,Klinisk kemi,Hälsouniversitetet
Liedberg, Bo (author)
Linköpings universitet,Tillämpad Fysik,Tekniska högskolan
Lundström, Ingemar (author)
Linköpings universitet,Tillämpad Fysik,Tekniska högskolan
Tengvall, Pentti (author)
Linköpings universitet,Tillämpad Fysik,Tekniska högskolan
show less...
 (creator_code:org_t)
SPIE-Intl Soc Optical Eng, 2000
2000
English.
In: Journal of Biomedical Optics. - : SPIE-Intl Soc Optical Eng. - 1083-3668 .- 1560-2281. ; 5:1, s. 51-55
  • Journal article (peer-reviewed)
Abstract Subject headings
Close  
  • A surface plasmon resonance (SPR) apparatus was used to investigate blood plasma coagulation in real time as a function of thromboplastin and heparin concentrations. The response curves were analyzed by curve fitting to a sigmoid curve equation, followed by extraction of the time constant. Clotting activation by thromboplastin resulted in increased time constant, as compared to spontaneously clotted plasma, in a dose dependent way. Addition of heparin to the thromboplastin-activated plasma counteracted this effect. Atomic force microscopy (AFM) pictures of sensor surfaces dried after completed clotting, revealed differences in fibrin network structures as a function of thromboplastin concentration, and the fiber thickness increased with decreased thromboplastin concentration. The physical reason for the SPR signal observed is ambiguous and is therefore discussed. However, the results summarized in the plots and the fibrin network properties observed by AFM correlate well with present common methods used to analyze blood coagulation.

Keyword

MEDICINE
MEDICIN

Publication and Content Type

ref (subject category)
art (subject category)

Find in a library

To the university's database

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view