Assessing the Accuracy of Continuous Glucose Monitoring (CGM) Calibrated With Capillary Values Using Capillary or Venous Glucose Levels as a Reference.

Andelin, M. (författare)
Department of Medicine, NU Hospital Group, Uddevalla, Sweden.
Kropff, J. (författare)
Department of Endocrinology, Academic Medical Center, University of Amsterdam, Amsterdam, Netherlands.
Matuleviciene, V. (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin,Institute of Medicine
visa fler...
Joseph, J.I. (författare)
Department of Anaesthesiology, Sidney Kimmel Medical College of Thomas Jefferson University, Philadelphia, PA, USA.
Attvall, Stig, 1950 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin,Institute of Medicine
Theodorsson, Elvar (författare)
Linköpings universitet,Avdelningen för mikrobiologi och molekylär medicin,Medicinska fakulteten,Region Östergötland, Klinisk kemi
Hirsch, I.B. (författare)
University of Washington, Seattle, WA, USA
Imberg, H. (författare)
Statistiska Konsultgruppen, Gothenburg, Sweden.
Dahlqvist, S. (författare)
Department of Medicine, NU Hospital Group, Uddevalla, Sweden
Klonoff, D. (författare)
Diabetes Research Institute, Mills-Peninsula Health Services, San Mateo, CA, USA.
Haraldsson, Börje, 1957 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin,Institute of Medicine
DeVries, J.H. (författare)
Department of Endocrinology, Academic Medical Center, University of Amsterdam, Amsterdam, Netherlands.
Lind, Marcus, 1976 (författare)
Gothenburg University,Göteborgs universitet,Institutionen för medicin,Institute of Medicine
visa färre...
Department of Medicine, NU Hospital Group, Uddevalla, Sweden Department of Endocrinology, Academic Medical Center, University of Amsterdam, Amsterdam, Netherlands. (creator_code:org_t)
2016-01-24
2016
Engelska.
Ingår i: Journal of Diabetes Science and Technology. - : Diabetes Technology Society. - 1932-2968. ; 10:4, s. 876-884
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: Using the standard venous reference for the evaluation of continuous glucose monitoring (CGM) systems could possibly negatively affect measured CGM accuracy since CGM are generally calibrated with capillary glucose and venous and capillary glucose concentrations differ. We therefore aimed to quantify the effect of using capillary versus venous glucose reference samples on estimated accuracy in capillary calibrated CGM.less thanbr /greater thanMethods: We evaluated 41 individuals with type 1 diabetes mellitus (T1DM) using the Dexcom G4 CGM system over 6 days. Patients calibrated their CGM devices with capillary glucose by means of the HemoCue system. During 2 visits, capillary and venous samples were simultaneously measured by HemoCue and compared to concomitantly obtained CGM readings. The mean absolute relative difference (MARD) was calculated using capillary and venous reference samples.less thanbr /greater thanResults: Venous glucose values were 0.83 mmol/L (15.0 mg/dl) lower than capillary values over all glycemic ranges, P less than .0001. Below 4 mmol/l (72 mg/dl), the difference was 1.25 mmol/l (22.5 mg/dl), P = .0001, at 4-10 mmol/l (72-180 mg/dl), 0.67 mmol/l (12.0 mg/dl), P less than .0001 and above 10 mmol/l (180 mg/dl), 0.95 mmol/l (17.1 mg/dl), P less than .0001. MARD was 11.7% using capillary values as reference compared to 13.7% using venous samples, P = .037. Below 4 mmol/l (72 mg/dl) MARD was 16.6% and 31.8%, P = .048, at 4-10 mmol/l (72-180 mg/dl) 12.1% and 12.6%, P = .32, above 10 mmol/l (180 mg/dl) 8.7% and 9.2%, P = .82.less thanbr /greater thanConclusion: Using capillary glucose concentrations as reference to evaluate the accuracy of CGM calibrated with capillary samples is associated with a lower MARD than using venous glucose as the reference. Capillary glucose concentrations were significantly higher than venous in all glycemic ranges.less thanbr /greater than (© 2016 Diabetes Technology Society.)

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Endokrinologi och diabetes (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Endocrinology and Diabetes (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Klinisk laboratoriemedicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Clinical Laboratory Medicine (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy