SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "L773:1932 2968 "

Sökning: L773:1932 2968

  • Resultat 1-10 av 62
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Börjesson, Anders, et al. (författare)
  • First Principles Studies of the Effect of Nickel Carbide Catalyst Composition on Carbon Nanotube Growth
  • 2010
  • Ingår i: The Journal of Physical Chemistry C. - : American Chemical Society. - 1932-7447 .- 1932-7455. ; 114:42, s. 18045-18050
  • Tidskriftsartikel (refereegranskat)abstract
    • Density functional theory calculations were used to investigate the stability of single-walled carbon nanotubes (CNTs) attached to nanoparticles. The total energies and the adhesion energies between the CNTs and the nanoparticles were calculated for systems where the nanoparticles were either pure Ni or Ni carbide. It was found that the adhesion between the CNT and a pure Ni cluster is stronger than between the same CNT and a Ni carbide cluster although the energy difference was small compared to the total adhesion energies. This adhesion strength implies that CNTs are likely to remain attached to both pure Ni and Ni carbide clusters and that either pure Ni or Ni carbide clusters may be docked onto the open CNT ends to achieve continued growth or electronic contacts between CNTs and electrode materials. The system with a CNT attached to a pure Ni cluster was found to be energetically favored compared to a system containing the same CNT attached to a Ni carbide. The difference in total energy implies that a CNT should act as a sink for C atoms dissolved in the Ni carbide cluster, which means that the dissolved C atoms will be drained from the cluster, yielding a pure metal in the zero Kelvin thermodynamic limit. It is argued that this draining procedure is likely to occur even if carbon is added to the cluster at a proper rate, for example, during CNT growth.
  •  
2.
  • Adolfsson, Peter, et al. (författare)
  • Automated Insulin Delivery Systems in Pediatric Type 1 Diabetes : A Narrative Review
  • 2024
  • Ingår i: Journal of Diabetes Science and Technology. - : Diabetes Technology Society. - 1932-2968.
  • Forskningsöversikt (refereegranskat)abstract
    • This narrative review assesses the use of automated insulin delivery (AID) systems in managing persons with type 1 diabetes (PWD) in the pediatric population. It outlines current research, the differences between various AID systems currently on the market and the challenges faced, and discusses potential opportunities for further advancements within this field. Furthermore, the narrative review includes various expert opinions on how different AID systems can be used in the event of challenges with rapidly changing insulin requirements. These include examples, such as during illness with increased or decreased insulin requirements and during physical activity of different intensities or durations. Case descriptions give examples of scenarios with added user-initiated actions depending on the type of AID system used. The authors also discuss how another AID system could have been used in these situations.
  •  
3.
  • Adolfsson, Peter, 1963, et al. (författare)
  • In-vitro performance of the Enlite sensor in various glucose concentrations during hypobaric and hyperbaric conditions
  • 2012
  • Ingår i: Journal of Diabetes Science and Technology. - Thousand Oaks, USA : Sage Publications. - 1932-2968. ; 6:6, s. 1375-1382
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: There is a need for reliable methods of glucose measurement in different environmental conditions. The objective of this in vitro study was to evaluate the performance of the Enlite® Sensor when connected to either the iPro™ Continuous Glucose Monitor recording device or the Guardian® REAL-Time transmitting device, in hypobaric and hyperbaric conditions.Methods: Sixteen sensors connected to eight iPro devices and eight Guardian REAL-Time devices were immersed in three beakers containing separate glucose concentrations: 52, 88, and 207 mg/dl (2.9, 4.9, and 11.3 mmol/liter). Two different pressure tests were conducted: a hypobaric test, corresponding to maximum 18000 ft/5500 m height, and a hyperbaric test, corresponding to maximum 100 ft/30 m depth. The linearity of the sensor signals in the different conditions was evaluated.Results: The sensors worked continuously, and the sensor signals were collected without interruption at all pressures tested. When comparing the input signals for glucose (ISIGs) and the different glucose concentrations during altered pressure, linearity (R(2)) of 0.98 was found. During the hypobaric test, significant differences (p < .005) were seen when comparing the ISIGs during varying pressure at two of the glucose concentrations (52 and 207 mg/dl), whereas no difference was seen at the 88 mg/dl glucose concentration. During the hyperbaric test, no differences were found.Conclusions: The Enlite Sensors connected to either the iPro or the Guardian REAL-Time device provided values continuously. In hyperbaric conditions, no significant differences were seen during changes in ambient pressure; however, during hypobaric conditions, the ISIG was significantly different in the low and high glucose concentrations.
  •  
4.
  • Adolfsson, Peter, 1963, et al. (författare)
  • The benefits of continuous glucose monitoring and a glucose monitoring schedule in individuals with type 1 diabetes during recreational diving
  • 2008
  • Ingår i: Journal of Diabetes Science and Technology. - : Diabetes Technology Society. - 1932-2968. ; 2:5, s. 778-784
  • Tidskriftsartikel (refereegranskat)abstract
    • BACKGROUND: Our objective is to evaluate the Medtronic CGMS continuous glucose monitoring system and plasma glucose (PG) measurement performed in a monitoring schedule as tools to identify individuals with type 1 diabetes at risk when diving.METHODS: We studied 24 adults, 12 type 1 diabetes subjects and 12 controls, during 5 recreational scuba dives performed on 3 consecutive days. The CGMS was used by all participants on all the days and all the dives. Comparisons were made between PG performed in a monitoring schedule during the days of diving, self-monitored blood glucose (SMBG) performed 2 weeks prior to diving, and the CGMS during the study.RESULTS: One hundred seventeen dives were performed. Hypoglycemia (<70 mg/dl) was found in six individuals and on nine occasions. However, no symptoms of hypoglycemia were present during or immediately postdiving. In one case, repetitive hypoglycemia prediving gave rise to a decision not to dive. None of the dives were aborted. The number of hypoglycemic episodes, 10 min prediving or immediately postdiving, were related to the duration of diabetes, r = 0.83 and p =0.01, and the percentage of SMBG values below target (<72 mg/dl), r = 0.65 and p =0.02. Moreover, the number of hypoglycemic episodes was also related to the total duration below low limit (<70 mg/dl), measured by the CGMS, r =0.74 and p =0.006.CONCLUSION: Safe dives are possible to achieve by well-informed, well-controlled individuals with type 1 diabetes. Using downloaded SMBG, CGMS, and repetitive PG in a monitoring schedule, it is possible to identify those subjects who are suitable for diving.
  •  
5.
  • Ahlén, Elsa, 1990, et al. (författare)
  • Glycemic control, renal complications, and current smoking in relation to excess risk of mortality in persons with type 1 diabetes
  • 2016
  • Ingår i: Journal of Diabetes Science and Technology. - : SAGE Publications. - 1932-2968. ; 10:5, s. 1006-1014
  • Tidskriftsartikel (refereegranskat)abstract
    • Abstract Background: A substantial excess risk of mortality still exists in persons with type 1 diabetes. The aim of this study was to evaluate the excess risk of mortality in persons with type 1 diabetes without renal complications who target goals for glycemic control and are nonsmokers. Furthermore, we evaluated risk factors of death due to hypoglycemia or ketoacidosis in young adults with type 1 diabetes. Methods: We evaluated a cohort based on 33 915 persons with type 1 diabetes and 169 249 randomly selected controls from the general population matched on age, sex, and county followed over a mean of 8.0 and 8.3 years, respectively. Hazard ratios (HRs) for all-cause and cardiovascular disease (CVD) mortality for persons with type 1 diabetes versus controls were estimated. Results: The adjusted HRs for all-cause and CVD mortality for persons with type 1 diabetes without renal complications (normoalbuminuria and eGFR ≥ 60 ml/min) and HbA1c ≤ 6.9% (52 mmol/mol) compared to controls were 1.22 (95% CI 0.98-1.52) and 1.03 (95% CI 0.66-1.60), respectively. The HRs increased with higher updated mean HbA1c. For nonsmokers in this group, the HRs for all-cause and CVD mortality were somewhat lower: 1.11 (95% CI 0.87-1.42) and 0.89 (95% CI 0.53-1.48) at updated mean HbA1c ≤ 6.9% (52 mmol/mol). HRs for significant predictors for deaths due to hypoglycemia or ketoacidosis in persons < 50 years were male sex 2.40 (95% CI 1.27-4.52), smoking 2.86 (95% CI 1.57-5.22), lower educational level 3.01 (95% CI 1.26-7.22), albuminuria or advanced kidney disease 2.83 (95% CI 1.63-4.93), earlier hospital diagnosis of hypoglycemia or ketoacidosis 2.30 (95% CI 1.20-4.42), and earlier diagnosis of intoxication 2.53 (95% CI 1.06-6.04). Conclusions: If currently recommended HbA1c targets can be reached, renal complications and smoking avoided in persons with type 1 diabetes, the excess risk of mortality will likely converge substantially to that of the general population.
  •  
6.
  • Andelin, M., et al. (författare)
  • Assessing the Accuracy of Continuous Glucose Monitoring (CGM) Calibrated With Capillary Values Using Capillary or Venous Glucose Levels as a Reference.
  • 2016
  • Ingår i: Journal of Diabetes Science and Technology. - : Diabetes Technology Society. - 1932-2968. ; 10:4, s. 876-884
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: Using the standard venous reference for the evaluation of continuous glucose monitoring (CGM) systems could possibly negatively affect measured CGM accuracy since CGM are generally calibrated with capillary glucose and venous and capillary glucose concentrations differ. We therefore aimed to quantify the effect of using capillary versus venous glucose reference samples on estimated accuracy in capillary calibrated CGM.less thanbr /greater thanMethods: We evaluated 41 individuals with type 1 diabetes mellitus (T1DM) using the Dexcom G4 CGM system over 6 days. Patients calibrated their CGM devices with capillary glucose by means of the HemoCue system. During 2 visits, capillary and venous samples were simultaneously measured by HemoCue and compared to concomitantly obtained CGM readings. The mean absolute relative difference (MARD) was calculated using capillary and venous reference samples.less thanbr /greater thanResults: Venous glucose values were 0.83 mmol/L (15.0 mg/dl) lower than capillary values over all glycemic ranges, P less than .0001. Below 4 mmol/l (72 mg/dl), the difference was 1.25 mmol/l (22.5 mg/dl), P = .0001, at 4-10 mmol/l (72-180 mg/dl), 0.67 mmol/l (12.0 mg/dl), P less than .0001 and above 10 mmol/l (180 mg/dl), 0.95 mmol/l (17.1 mg/dl), P less than .0001. MARD was 11.7% using capillary values as reference compared to 13.7% using venous samples, P = .037. Below 4 mmol/l (72 mg/dl) MARD was 16.6% and 31.8%, P = .048, at 4-10 mmol/l (72-180 mg/dl) 12.1% and 12.6%, P = .32, above 10 mmol/l (180 mg/dl) 8.7% and 9.2%, P = .82.less thanbr /greater thanConclusion: Using capillary glucose concentrations as reference to evaluate the accuracy of CGM calibrated with capillary samples is associated with a lower MARD than using venous glucose as the reference. Capillary glucose concentrations were significantly higher than venous in all glycemic ranges.less thanbr /greater than (© 2016 Diabetes Technology Society.)
  •  
7.
  • Anderson, J., et al. (författare)
  • Effect on glycemic control by short- and long-term use of continuous glucose monitoring in clinical practice
  • 2011
  • Ingår i: Journal of diabetes science and technology. - : SAGE Publications. - 1932-2968. ; 5:6, s. 1472-9
  • Tidskriftsartikel (refereegranskat)abstract
    • BACKGROUND: In Sweden, patients with diabetes mellitus frequently receive short-term (<3 months) continuous glucose monitoring (CGM) to study glucose patterns or long-term CGM to treat poor glycemic control or severe hypoglycemia. The effects of CGM on glycemic control in clinical practice in relation to indication and duration of use has not been completely studied. METHODS: Patients with diabetes, among which 99% were diagnosed as type 1, receiving CGM at 10 outpatient clinics in Sweden were studied retrospectively. Long-term use of CGM was defined as >/= 3 months use of CGM and short-term as <3 months. A control group matched on start date and date of latest value 3 months after the start was selected for both long- and short-term groups. RESULTS: In 34 long-term users of CGM, over a mean follow-up of 1.1 years, the adjusted mean difference of hemoglobin A1c (HbA1c) compared with controls (n = 408) was -0.76 (95% confidence interval -1.17; -0.33, p < .001). Long-term users with indications for high HbA1c (n = 15) had a reduction of 1.2% in HbA1c from 10.1 to 8.9% (p = .003), whereas patients with hypoglycemia as their indication (n = 16) decreased by 0.3% (p = .17). Nonsevere hypoglycemic events decreased in long-term users within the same follow-up period (p = .004). Short-term users showed no statistically significant improvement in HbA1c compared with controls at 1.1 years (n = 41), p = .85 or at 2.6 years (n = 43), p = .19. CONCLUSION: Long-term CGM use was associated with improved glycemic control in clinical practice and a reduction in nonsevere hypoglycemic events, whereas short-term use had no effect on HbA1c. The effect on glycemic control varied by indication.
  •  
8.
  • Barnard, Katharine, et al. (författare)
  • Impact of Chronic Sleep Disturbance for People Living With T1 Diabetes.
  • 2016
  • Ingår i: Journal of diabetes science and technology. - : SAGE Publications. - 1932-2968. ; 10:3, s. 762-767
  • Tidskriftsartikel (refereegranskat)abstract
    • The aim was to explore personal experiences and to determine the impact of impaired sleep on well-being and diabetes-related activities/decision making among a cohort of people living with T1D.
  •  
9.
  • Barnard, Katharine, et al. (författare)
  • Impact of Type 1 Diabetes Technology on Family Members/Significant Others of People With Diabetes.
  • 2016
  • Ingår i: Journal of diabetes science and technology. - : SAGE Publications. - 1932-2968. ; 10:4, s. 824-30
  • Tidskriftsartikel (refereegranskat)abstract
    • The aim was to explore the impact of diabetes-related technology to ensure that such devices are used in a way that returns maximum benefit from a medical and psychological perspective.Spouses and caregivers of people with type 1 diabetes were invited to complete an online questionnaire about their experiences with diabetes technologies used by their family members. Participants were recruited via the Glu online community website. Questions explored impact on daily living, frequency and severity of hypoglycemia, and diabetes-related distress.In all, 100 parents/caregivers and 74 partners participated in this survey. Average (mean) duration of living with a person with type 1 diabetes was 16 years (SD = 13) for partners, with duration of diabetes for children being 4.2 ± 3.2 years. Average duration of current therapy was 8.3 ± 7.3 years for adults and 3.4 ± 2.9 years for children. Of the participants, 86% partners and 82% parents/caregivers reported diabetes technology had made it easier for their family members to achieve blood glucose targets. Compared to partners, parents/caregivers reported more negative emotions (P < .001) and decreased well-being (P < .001) related to their family members type 1 diabetes. Diabetes-related distress was common, as was sleep disturbance associated with device alarms and fear of hypoglycemia. Reduced frequency and severity of hypoglycemia related to device use was reported by approximately half of participants.There is little doubt about the medical benefit of diabetes technologies and their uptake is increasing but some downsides were reported. Barriers to uptake of technologies lie beyond the mechanics of diabetes management. Supporting users in using diabetes technology to achieve the best possible glycemic control, in the context of their own life, is crucial. Furthermore, understanding these issues with input from the type 1 diabetes community including family members and caregivers will help innovation and design of new technology.
  •  
10.
  • Battelino, T, et al. (författare)
  • Guideline Development for Medical Device Technology: Issues for Consideration
  • 2023
  • Ingår i: Journal of diabetes science and technology. - : SAGE Publications. - 1932-2968. ; 17:6, s. 1698-1710
  • Tidskriftsartikel (refereegranskat)abstract
    • Advances in the development of innovative medical devices and telehealth technologies create the potential to improve the quality and efficiency of diabetes care through collecting, aggregating, and interpreting relevant health data in ways that facilitate more informed decisions among all stakeholder groups. Although many medical societies publish guidelines for utilizing these technologies in clinical practice, we believe that the methodologies used for the selection and grading of the evidence should be revised. In this article, we discuss the strengths and limitations of the various types of research commonly used for evidence selection and grading and present recommendations for modifying the process to more effectively address the rapid pace of device and technology innovation and new product development.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-10 av 62
Typ av publikation
tidskriftsartikel (55)
forskningsöversikt (5)
konferensbidrag (2)
Typ av innehåll
refereegranskat (58)
övrigt vetenskapligt/konstnärligt (4)
Författare/redaktör
Jendle, Johan, 1963- (16)
Adolfsson, Peter, 19 ... (9)
Lind, Marcus, 1976 (9)
Dahlqvist, S. (5)
Liska, J (4)
Mowitz, Martin (3)
visa fler...
Johansson, Rolf (2)
Garcia-Gabin, Winsto ... (2)
Hellman, Jarl (2)
Thomas, Andreas (2)
Pivodic, A. (2)
Kosiborod, M. (2)
Wang, J. (1)
Rydén, L. (1)
Green, J. (1)
Svensson, Ann-Marie, ... (1)
Östlundh, Linda, 197 ... (1)
Theodorsson, Elvar (1)
Berg, L (1)
Li, C. (1)
Cosentino, F (1)
Simrén, Magnus, 1966 (1)
Anderson, J (1)
Roxhed, Niclas (1)
Brismar, K (1)
Adami, HO (1)
Ekelund, M. (1)
Adamson, U (1)
Gudbjörnsdottir, Sof ... (1)
Rosengren, Annika, 1 ... (1)
Adolfsson, Peter (1)
Hanas, Ragnar (1)
Zaharieva, Dessi P. (1)
Dovc, Klemen (1)
Ornhagen, Hans (1)
Eriksson, Bengt M. (1)
Hanås, Ragnar, 1951 (1)
Hartvig, N. V. (1)
Kaas, A. (1)
Gautham, Raghavendha ... (1)
Örnhagen, Hans (1)
Anderberg, Peter (1)
Bruze, Magnus (1)
Russell, S J (1)
Wernerman, J (1)
Agvall, Björn (1)
Galozy, Alexander, 1 ... (1)
Bellander, Bo Michae ... (1)
Ahlén, Elsa, 1990 (1)
Wedel, Hans (1)
visa färre...
Lärosäte
Örebro universitet (21)
Göteborgs universitet (20)
Karolinska Institutet (16)
Uppsala universitet (6)
Lunds universitet (6)
Linköpings universitet (4)
visa fler...
Kungliga Tekniska Högskolan (3)
Chalmers tekniska högskola (3)
Högskolan i Halmstad (1)
Mälardalens universitet (1)
Jönköping University (1)
Malmö universitet (1)
Högskolan i Borås (1)
Blekinge Tekniska Högskola (1)
visa färre...
Språk
Engelska (62)
Forskningsämne (UKÄ/SCB)
Medicin och hälsovetenskap (43)
Teknik (5)
Naturvetenskap (3)

År

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy