- Holmer, Mattias, et al.
(författare)
-
On-line Heart Rate Monitoring Using the Extra-corporeal Pressure Sensors of a Dialysis Machine
- 2012
-
Ingår i: Biomedical engineering – 2012 : proceedings of international conference. - 2029-3380.
-
Konferensbidrag (refereegranskat)abstract
- Heart rate can be extracted from the extracorporeal venous pressure signal of a dialysis machine. The results are of comparable accuracy and reliability to the ones obtained by the PPG reference signal from a pulse oximeter. Difficulties that occur during heart rate estimation were determined, and some can be overcome by a slight adjustment of blood flow rate. The described techniques, after being implemented into the dialysis machines, would help to improve current hemodialysis safety throughout the treatment by on-line monitoring of the cardiac activity.
|
|