SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "L773:9781479960972 "

Sökning: L773:9781479960972

  • Resultat 1-3 av 3
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Allahgholi, A., et al. (författare)
  • AGIPD 1.0 : The high-speed high dynamic range readout ASIC for the adaptive gain integrating pixel detector at the European XFEL
  • 2014
  • Ingår i: 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2014. - : Institute of Electrical and Electronics Engineers (IEEE). - 9781479960972
  • Konferensbidrag (refereegranskat)abstract
    • AGIPD is a hybrid pixel X-ray detector developed by a collaboration between Deutsches Elektronen-Synchrotron (DESY), Paul-Scherrer-Institute (PSI), University of Hamburg and the University of Bonn. The detector is designed to comply with the requirements of the European XFEL. The radiation tolerant Application Specific Integrated Circuit (ASIC) is designed with the following highlights: high dynamic range, spanning from single photon sensitivity up to 104 × 12.4 keV photons, achieved by the use of dynamic gain switching, auto-selecting one of 3 gains of the charge sensitive pre-amplifier. To cope with the unique features of the European XFEL source, image data is stored in 352 analogue memory cells per pixel. The selected gain is stored in the same way and depth, encoded as one of 3 voltage levels. These memories are operated in random-access mode at 4.5MHz frame rate. Data is read out on a row-by-row basis via multiplexers to the DAQ system for digitisation during the 99.4ms gap between the bunch trains of the European XFEL. The AGIPD 1.0 ASIC features 64×64 pixels with a pixel area of 200×200 μm2. It is bump-bonded to a 500 μm thick silicon sensor. The principles of the chip architecture were proven in different experiments and the ASIC characterization was performed with a series of development prototypes. The mechanical concept of the detector system was developed in close contact with the XFEL beamline scientists to ensure a seamless integration into the beamline setup and is currently being manufactured. The first single module system was successfully tested at APS1 the high dynamic range allows imaging of the direct synchrotron beam along with single photon sensitivity and burst imaging of 352 subsequent frames synchronized to the source.
  •  
2.
  • Dahlbom, Magnus, et al. (författare)
  • Effects of high photon fluence rate from therapeutic radionuclides on preclinical and clinical PET systems
  • 2016
  • Ingår i: 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2014. - 9781479960972
  • Konferensbidrag (refereegranskat)abstract
    • Tumor response in radionuclide therapy can be monitored with PET/CT and/or PET/MR. A high background photon fluence from a therapy radionuclide may influence both image quality and quantification, when imaging is performed intra-therapeutically, i.e. with high activity of the therapeutic radionuclide present. Here, count losses and image distortion have been investigated for preclinical and clinical PET systems with different detector designs. The effect on the spatial resolution was studied with a point source of 22Na in a background of 99mTc, where 99mTc emulated the photon emission from a therapeutic radionuclide. An in-house made mouse phantom with silicon tubes filled with 99mTc with a centrally placed 22Na point source was used. For the clinical systems, a 70 cm long NEMA PET Scatter Phantom was used, with a 22Na point source placed at the center whereas the off-center silicon tube was filled with 99mTc. In addition, image quality was also evaluated in the presence of different levels of 99mTc with a 18F-filled NEMA image quality phantom on the preclinical systems and a 18F-filled Jaszczak phantom on the clinical system. Preclinical PET systems with different detector geometries showed that the addition of 99mTc affected the count rate capability considerably, especially those with a low number of read-out channels. The coincidence rate for was significantly reduced when high activities of 99mTc were present. The clinical PET system also showed an effect of reduced coincidence rate with increased photon fluence rate. At high 99mTc activities, the spatial resolution was degraded for both the preclinical and the clinical systems. The quantitative capability of PET systems used intra-therapeutically is significantly affected by the additional high photon fluence rate. The dead-time correction implemented on some of the investigated PET systems, was able to accurately compensate for the coincidence count losses. The reduced spatial resolution at high photon fluence rate, however, remains a potentially limiting factor.
  •  
3.
  • Ljungberg, Michael, et al. (författare)
  • Monte Carlo simulations of the GE discovery alcyone CZT SPECT system
  • 2016
  • Ingår i: 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2014. - 9781479960972
  • Konferensbidrag (refereegranskat)abstract
    • Compact SPECT systems with cadmium zinc telluride (CZT) solid-state detectors with improved energy resolution and shorter acquisition times have recently been introduced. These systems have, however, different energy characteristics compared to NaI(Tl) crystal-based cameras. There is therefore a need to develop new simulation models for these cameras. We modeled the charge transport within the CZT detectors for a GE Discovery 530c/570c SPECT system with multiple pinhole collimators employing the SIMIND Monte Carlo program and validated simulations against measurements. The incomplete charge collection between the anode and cathode in the pixilated CZT was modeled with the Hecht equation. The simulation also included charge-sharing effects across pixels due to physical interactions and charge diffusion. To validate our CZT-model 99mTc and 123I point sources and a 201Tl line source were acquired and measured energy spectra were compared with simulated energy spectra. The Monte Carlo simulated energy spectra agreed well with the experimental measurements within the photopeak, overestimated the k-edge x-ray escape peaks of Cd and Te, and slightly underestimated the remainder of the tail. Comparisons for a cardiac insert with a defect in an elliptical Data Spectrum phantom were also performed. Here, simulated projections were read into the GE Xeleris system for reconstruction. We found good agreement in image reconstruction visually. We conclude that it is feasible to simulate CZT detectors with good agreement using the SIMIND code.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-3 av 3

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy