  • Citerio, Giuseppe, et al. (författare)
  • Multicenter clinical assessment of the raumedic Neurovent-P intracranial pressure sensor : A report by the brainIT group
  • 2008
  • Ingår i: Neurosurgery. - 0148-396X .- 1524-4040. ; 63:6, s. 1152-1158
  • Tidskriftsartikel (refereegranskat)abstract
    • OBJECTIVE: The aim of this study was to evaluate the robustness and zero-drift of an intracranial pressure sensor, Neurovent-P (Raumedic AG, Munchberg, Germany), when used in the clinical environment. METHODS: A prospective multicenter trial, conforming to the International Organization for Standardization 14155 Standard, was conducted in 6 European BrainIT centers between July 2005 and December 2006. Ninety-nine catheters were used. The study was observational, followed by a centralized sensor bench test after catheter removal. RESULTS: The mean recorded value before probe insertion was 0.17 +/- 1.1 mm Hg. Readings outside the range 1 mm Hg were recorded in only 3 centers on a total of 15 catheters. Complications were minimal and mainly related to the insertion bolt. The mean recorded pressure value at removal was 0.8 +/- 2.2 mm Hg. No relationship was identified between postremoval reading and length of monitoring. The postremoval bench test indicated the probability of a system failure, defined as a drift of more than 3 mm Hg, at a range between 12 and 17%. CONCLUSION: The Neurovent-P catheter performed well in clinical use in terms of robustness. The majority of technical complications were associated with the bolt fixation technology. Adverse events were rare and clinically nonsignificant. Despite the earlier reported excellent bench test zero-drift rates, under the more demanding clinical conditions, zero-drift rate remains a concern with catheter tip strain gauge technology. This performance is similar, and not superior, to other intracranial pressure devices.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-2 av 2

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy