SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "WFRF:(Dahlgren Christina Vallhagen) "

Sökning: WFRF:(Dahlgren Christina Vallhagen)

  • Resultat 1-6 av 6
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  •  
2.
  • Andersson, Karin M., 1989-, et al. (författare)
  • Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning
  • 2014
  • Ingår i: Journal of Applied Clinical Medical Physics. - : John Wiley & Sons. - 1526-9914. ; 15:5, s. 112-119
  • Tidskriftsartikel (refereegranskat)abstract
    • Metal objects in the body such as hip prostheses cause artifacts in CT images. When CT images degraded by artifacts are used for treatment planning of radiotherapy, the artifacts can yield inaccurate dose calculations and, for particle beams, erroneous penetration depths. A metal artifact reduction software (O-MAR) installed on a Philips Brilliance Big Bore CT has been tested for applications in treatment planning of proton radiotherapy. Hip prostheses mounted in a water phantom were used as test objects. Images without metal objects were acquired and used as reference data for the analysis of artifact-affected regions outside of the metal objects in both the O-MAR corrected and the uncorrected images. Water equivalent thicknesses (WET) based on proton stopping power data were calculated to quantify differences in the calculated proton beam penetration for the different image sets. The WET to a selected point of interest between the hip prostheses was calculated for several beam directions of clinical relevance. The results show that the calculated differences in WET relative to the reference case were decreased when the O-MAR algorithm was applied. WET differences up to 2.0 cm were seen in the uncorrected case while, for the O-MAR corrected case, the maximum difference was decreased to 0.4 cm. The O-MAR algorithm can significantly improve the accuracy in proton range calculations. However, there are some residual effects, and the use of proton beam directions along artifact streaks should only be used with caution and appropriate margins.
  •  
3.
  • Andersson, Karin M., 1989-, et al. (författare)
  • Evaluation of two commercial CT metal artifact reduction algorithms for use in proton radiotherapy treatment planning in the head and neck area
  • 2018
  • Ingår i: Medical physics (Lancaster). - : Wiley-Blackwell Publishing Inc.. - 0094-2405 .- 2473-4209. ; 45:10, s. 4329-4344
  • Tidskriftsartikel (refereegranskat)abstract
    • PURPOSE: To evaluate two commercial CT metal artifact reduction (MAR) algorithms for use in proton treatment planning in the head and neck (H&N) area.METHODS: An anthropomorphic head phantom with removable metallic implants (dental fillings or neck implant) was CT-scanned to evaluate the O-MAR (Philips) and the iMAR (Siemens) algorithms. Reference images were acquired without any metallic implants in place. Water equivalent thickness (WET) was calculated for different path directions and compared between image sets. Images were also evaluated for use in proton treatment planning for parotid, tonsil, tongue base, and neck node targets. The beams were arranged so as to not traverse any metal prior to the target, enabling evaluation of the impact on dose calculation accuracy from artifacts surrounding the metal volume. Plans were compared based on γ analysis (1 mm distance-to-agreement/1% difference in local dose) and dose volume histogram metrics for targets and organs at risk (OARs). Visual grading evaluation of 30 dental implant patient MAR images was performed by three radiation oncologists.RESULTS: In the dental fillings images, ΔWET along a low-density streak was reduced from -17.0 to -4.3 mm with O-MAR and from -16.1 mm to -2.3 mm with iMAR, while for other directions the deviations were increased or approximately unchanged when the MAR algorithms were used. For the neck implant images, ΔWET was generally reduced with MAR but residual deviations remained (of up to -2.3 mm with O-MAR and of up to -1.5 mm with iMAR). The γ analysis comparing proton dose distributions for uncorrected/MAR plans and corresponding reference plans showed passing rates >98% of the voxels for all phantom plans. However, substantial dose differences were seen in areas of most severe artifacts (γ passing rates of down to 89% for some cases). MAR reduced the deviations in some cases, but not for all plans. For a single patient case dosimetrically evaluated, minor dose differences were seen between the uncorrected and MAR plans (γ passing rate approximately 97%). The visual grading of patient images showed that MAR significantly improved image quality (P < 0.001).CONCLUSIONS: O-MAR and iMAR significantly improved image quality in terms of anatomical visualization for target and OAR delineation in dental implant patient images. WET calculations along several directions, all outside the metallic regions, showed that both uncorrected and MAR images contained metal artifacts which could potentially lead to unacceptable errors in proton treatment planning. ΔWET was reduced by MAR in some areas, while increased or unchanged deviations were seen for other path directions. The proton treatment plans created for the phantom images showed overall acceptable dose distributions differences when compared to the reference cases, both for the uncorrected and MAR images. However, substantial dose distribution differences in the areas of most severe artifacts were seen for some plans, which were reduced by MAR in some cases but not all. In conclusion, MAR could be beneficial to use for proton treatment planning; however, case-by-case evaluations of the metal artifact-degraded images are always recommended.
  •  
4.
  • Dahlgren, Christina Vallhagen, et al. (författare)
  • Portal dose image verification : the collapsed cone superposition method applied with different electronic portal imaging devices.
  • 2006
  • Ingår i: Physics in Medicine and Biology. - 0031-9155. ; 51:2, s. 335-49
  • Tidskriftsartikel (refereegranskat)abstract
    • Two different commercial electronic portal imaging devices (EPIDs), one based on a liquid ion chamber matrix and the other based on a fluoroscopic CCD camera, were used to acquire readings that, through a calibration procedure, provided images proportional to the absolute dose to a virtual water slab located at the EPID plane. The transformation of the matrix ion chamber image into a portal dose image (PDI) was based on a published relationship between dose rate and ionization current. For the fluoroscopic CCD-camera-based system, the transformation was based on a deconvolution with a radial light scatter kernel. Local response variations were corrected in the images from both systems using open field fluence maps. The acquired PDIs were compared with PDIs calculated with the collapsed cone superposition method for a three-dimensional detector model in water equivalent buildup material. The calculation model was based on the beam modelling and geometrical description of the treatment unit and energy used for treatment planning in a kernel-based system. The validity of the calculation method was evaluated for several field shapes and thicknesses of patient phantoms for the matrix ion chamber at 6 MV x-rays and for the camera-based EPID at 6 and 15 MV x-rays. The agreement between predicted and measured PDIs was evaluated with dose comparisons at points of interest and gamma index calculations. The average area failing the passing criteria in dose and position deviation was analysed to validate the performance of the method. For the matrix ion chamber on average an area less than 1% fails the passing criteria of 3 mm and 3%. For the camera-based EPID, the average area is 7% and 1% for 6 and 15 MV, respectively. The overall agreement centrally in the fields was 0.1 +/- 1.6% (1 sd) for the camera-based EPID and -0.1 +/- 1.6% (1 sd) for the matrix ion chamber. Thus, an absolute dose calibrated EPID could validate the delivered dose to the patient by comparing a calculated and a measured PDI.
  •  
5.
  • Edvardsson, Anneli, et al. (författare)
  • Robustness and dosimetric verification of hippocampal-sparing craniospinal pencil beam scanning proton plans for pediatric medulloblastoma
  • 2024
  • Ingår i: Physics and Imaging in Radiation Oncology. - : Elsevier. - 2405-6316. ; 29
  • Tidskriftsartikel (refereegranskat)abstract
    • Background and PurposeHippocampal-sparing (HS) is a method that can potentially reduce late cognitive complications for pediatric medulloblastoma (MB) patients treated with craniospinal proton therapy (PT). The aim of this study was to investigate robustness and dosimetric plan verification of pencil beam scanning HS PT.Materials and MethodsHS and non-HS PT plans for the whole brain part of craniospinal treatment were created for 15 pediatric MB patients. A robust evaluation of the plans was performed. Plans were recalculated in a water phantom and measured field-by-field using an ion chamber detector at depths corresponding to the central part of hippocampi. All HS and non-HS fields were measured with the standard resolution of the detector and in addition 16 HS fields were measured with high resolution. Measured and planned dose distributions were compared using gamma evaluation.ResultsThe median mean hippocampus dose was reduced from 22.9 Gy (RBE) to 8.9 Gy (RBE), while keeping CTV V95% above 95 % for all nominal HS plans. HS plans were relatively robust regarding hippocampus mean dose, however, less robust regarding target coverage and maximum dose compared to non-HS plans. For standard resolution measurements, median pass rates were 99.7 % for HS and 99.5 % for non-HS plans (p < 0.001). For high-resolution measurements, median pass rates were 100 % in the hippocampus region and 98.2 % in the surrounding region.ConclusionsA substantial reduction of dose in the hippocampus region appeared feasible. Dosimetric accuracy of HS plans was comparable to non-HS plans and agreed well with planned dose distribution in the hippocampus region.
  •  
6.
  • La Tessa, C., et al. (författare)
  • Out-of-field dose studies with an anthropomorphic phantom : Comparison of X-rays and particle therapy treatments
  • 2012
  • Ingår i: Radiotherapy and Oncology. - : Elsevier BV. - 0167-8140 .- 1879-0887. ; 105:1, s. 133-138
  • Tidskriftsartikel (refereegranskat)abstract
    • Background and purpose: Characterization of the out-of-field dose profile following irradiation of the target with a 3D treatment plan delivered with modern techniques. Methods: An anthropomorphic RANDO phantom was irradiated with a treatment plan designed for a simulated 5×2×5cm 3 tumor volume located in the center of the head. The experiment was repeated with all most common radiation treatment types (photons, protons and carbon ions) and delivery techniques (Intensity Modulated Radiation Therapy, passive modulation and spot scanning). The measurements were performed with active diamond detector and passive thermoluminescence (TLD) detectors to investigate the out-of-field dose both inside and outside the phantom. Results: The highest out-of-field dose values both on the surface and inside the phantom were measured during the treatment with 25 MV photons. In the proximity of the Planned Target Volume (PTV), the lowest lateral dose profile was observed for passively modulated protons mainly because of the presence of the collimator in combination with the chosen volume shape. In the far out-of-field region (above 100 mm from the PTV), passively modulated ions were characterized by a less pronounced dose fall-off in comparison with scanned beams. Overall, the treatment with scanned carbon ions delivered the lowest dose outside the target volume. Conclusions: For the selected PTV, the use of the collimator in proton therapy drastically reduced the dose deposited by ions or photons nearby the tumor. Scanning modulation represents the optimal technique for achieving the highest dose reduction far-out-of-field.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-6 av 6

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy