- Ljunggren, Kaj, et al.
(författare)
-
Beta camera low activity tumor imaging
- 1993
-
Ingår i: Acta Oncologica. - : Informa UK Limited. - 1651-226X .- 0284-186X. ; 32:7-8, s. 869-872
-
Tidskriftsartikel (refereegranskat)abstract
- A new technique, the beta camera, to complement film autoradiography, with fast quantitative imaging of beta particle-emitting radionuclides has been developed. It consists of a thin plastic scintillator and a light-sensitive microchannel plate detector. The thin tissue sample is mounted on the scintillator. Our first system had a high background and a moderate spatial resolution of 900 microns. We now report an improved system with a photomultiplier tube mounted on the scintillator of the microchannel plate detector. Only events registered by both detectors are accepted. A fast coincidence unit processes the signals, and if a time overlap exists, an event is generated in the beta camera. In the coincidence mode, images with low activity distribution of 201Tl (count rate 1 s-1) in 50 microns-thick slices of a human glioma tumor could be recorded with a spatial resolution of 500 microns.
|
|