SwePub
Tyck till om SwePub Sök här!
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "WFRF:(Fredenberg Erik PhD 1979 ) "

Sökning: WFRF:(Fredenberg Erik PhD 1979 )

  • Resultat 1-10 av 29
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Cederström, Björn, et al. (författare)
  • Observer-model optimization of X-ray system in photon-counting breast imaging
  • 2011
  • Ingår i: Nuclear Instruments and Methods in Physics Research Section A. - : Elsevier BV. - 0168-9002 .- 1872-9576. ; 648:Supplement 1, s. S54-S57
  • Tidskriftsartikel (refereegranskat)abstract
    • An ideal-observer model is applied to optimize the design of an X-ray tube intended for use in a multi-slit scanning photon-counting mammography system. The design is such that the anode and the heel effect are reversed and the projected focal spot is smallest at the chest wall. Using linear systems theory, detectability and dose efficiency for a 0.1-mm disk are calculated for different focal spot sizes and anode angles. It is shown that the image acquisition time can be reduced by about 25% with spatial resolution and dose efficiency improved near the chest wall and worsened further away. The image quality is significantly more homogeneous than for the conventional anode orientation, both with respect to noise and detectability of a small object. With the tube rotated 90°, dose efficiency can be improved by 20% for a fixed image acquisition time. 
  •  
2.
  • Cederström, Björn, et al. (författare)
  • The influence of anatomical noise on optimal beam quality in mammography
  • 2014
  • Ingår i: Medical physics (Lancaster). - : Wiley. - 0094-2405. ; 41:12, s. 121903-
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: Beam-quality optimization in digital mammography traditionally considers detection of a target obscured by quantum noise in a homogeneous background. This does not correspond well to the clinical imaging task because real mammographic images contain a complex superposition of anatomical structures, resulting in anatomical noise that may dominate over quantum noise. The purpose of this paper is to assess the influence on optimal beam quality in mammography when anatomical noise is taken into account. Methods: The detectability of microcalcifications and masses was quantified using a theoretical ideal-observer model that included quantum noise as well as anatomical noise and a simplified model of a photon-counting mammography system. The outcome was experimentally verified using two types of simulated tissue phantoms. Results: The theoretical model showed that the detectability of tumors and microcalcifications behaves differently with respect to beam quality and dose. The results for small microcalcifications were similar to what traditional optimization methods yield, which is to be expected because quantum noise dominates over anatomical noise at high spatial frequencies. For larger tumors, however, low-frequency anatomical noise was the limiting factor. Because anatomical structure noise has similar energy dependence as tumor contrast, the optimal x-ray energy was found to be higher and the useful energy region was wider than traditional methods suggest. A simplified scalar model was able to capture this behavior using a fitted noise mixing parameter. The phantom measurements confirmed these theoretical results. Conclusions: It was shown that since quantum noise constitutes only a small fraction of the noise, the dose could be reduced substantially without sacrificing tumor detectability. Furthermore, when anatomical noise is included, the tube voltage can be increased well beyond what is conventionally considered optimal and used clinically, without loss of image quality. However, no such conclusions can be drawn for the more complex mammographic imaging task as a whole. (C) 2014 American Association of Physicists in Medicine.
  •  
3.
  • Dahlman, Nils, et al. (författare)
  • Evaluation of photon-counting spectral breast tomosynthesis
  • 2011
  • Ingår i: Medical Imaging 2011. - : SPIE.
  • Konferensbidrag (refereegranskat)abstract
    • We have designed a mammography system that for the first time combines photon-counting spectral imaging with tomosynthesis. The present study is a comprehensive physical evaluation of the system; tomosynthesis, spectral imaging, and the combination of both are compared using an ideal-observer model that takes anatomical noise into account. Predictions of signal and noise transfer through the system are verified by contrast measurements on a tissue phantom and 3D measurements of MTF and NPS. Clinical images acquired with the system are discussed in view of the model predictions.
  •  
4.
  • Fredenberg, Erik, PhD, 1979-, et al. (författare)
  • A low-absorption x-ray energy filter for small-scale applications
  • 2009
  • Ingår i: Optics Express. - : The Optical Society. - 1094-4087. ; 17:14, s. 11388-11398
  • Tidskriftsartikel (refereegranskat)abstract
    • We present an experimental and theoretical evaluation of an x-ray energy filter based on the chromatic properties of a prism-array lens (PAL). It is intended for small-scale applications such as medical imaging. The PAL approximates a Fresnel lens and allows for high efficiency compared to filters based on ordinary refractive lenses, however at the cost of a lower energy resolution. Geometrical optics was found to provide a good approximation for the performance of a flawless lens, but a field-propagation model was used for quantitative predictions. The model predicted a 0.29 ΔE/E energy resolution and an intensity gain of 6.5 for a silicon PAL at 23.5 keV. Measurements with an x-ray tube showed good agreement with the model in energy resolution and peak energy, but a blurred focal line contributed to a 29% gain reduction. We believe the blurring to be caused mainly by lens imperfections, in particular at the periphery of the lens.
  •  
5.
  •  
6.
  • Fredenberg, Erik, PhD, 1979-, et al. (författare)
  • A photon-counting detector for dual-energy breast tomosynthesis
  • 2009
  • Ingår i: Medical Imaging 2009. - : SPIE. ; , s. 72581-
  • Konferensbidrag (refereegranskat)abstract
    • We present the first evaluation of a recently developed silicon-strip detector for photon-counting dual-energy breast tomosynthesis. The detector is well suited for tomosynthesis with high dose efficiency and intrinsic scatter rejection. A method was developed for measuring the spatial resolution of a system based on the detector in terms of the three-dimensional modulation transfer function (MTF). The measurements agreed well with theoretical expectations, and it was seen that depth resolution was won at the cost of a slightly decreased lateral resolution. This may be a justifiable trade-off as clinical images acquired with the system indicate improved conspicuity of breast lesions. The photon-counting detector enables dual-energy subtraction imaging with electronic spectrumsplitting. This improved the detectability of iodine in phantom measurements, and the detector was found to be stable over typical clinical acquisition times. A model of the energy resolution showed that further improvements are witn reach by optimization of the detector.
  •  
7.
  • Fredenberg, Erik, PhD, 1979-, et al. (författare)
  • A Tunable Energy Filter for Medical X-Ray Imaging
  • 2008
  • Ingår i: X-Ray Optics and Instrumentation. - : Hindawi. - 1687-7632 .- 1687-7640. ; 2008
  • Tidskriftsartikel (refereegranskat)abstract
    • A multiprism lens (MPL) is a refractive X-ray lens, and its chromatic properties can be employed in an energy filtering setup to obtain a narrow tunable X-ray spectrum. We present the first evaluation of such a filter for medical X-ray imaging. The experimental setup yields a 6.6 gain of flux at 20 keV, and we demonstrate tunability by altering the energy spectrum to center also around 17 and 23 keV. All measurements are found to agree well with ray-tracing and a proposed geometrical model. Compared to a model mammography system with absorption filtering, the experimental MPL filter reduces dose 13–25% for 3–7 cm breasts if the spectrum is centered around the optimal energy. Additionally, the resolution is improved 2.5 times for a 5 cm breast. The scan time is increased 3 times but can be reduced with a slightly decreased energy filtering and resolution.
  •  
8.
  • Fredenberg, Erik, PhD, 1979-, et al. (författare)
  • Contrast-enhanced dual-energy subtraction imaging using electronic spectrum-splitting and multi-prism x-ray lenses
  • 2008
  • Ingår i: Medical Imaging 2008 - Physics of Medical Imaging. - San Diego, CA, USA : SPIE. - 9780819470973 ; , s. 91310-91310
  • Konferensbidrag (refereegranskat)abstract
    • Dual-energy subtraction imaging (DES) is a method to improve the detectability of contrast agents over a lumpy background. Two images, acquired at x-ray energies above and below an absorption edge of the agent material, are logarithmically subtracted, resulting in suppression of the signal from the tissue background and a relative enhancement of the signal from the agent. Although promising, DES is still not widely used in clinical practice. One reason may be the need for two distinctly separated x-ray spectra that are still close to the absorption edge, realized through dual exposures which may introduce motion unsharpness. In this study, electronic spectrum-splitting with a silicon-strip detector is theoretically and experimentally investigated for a mammography model with iodinated contrast agent. Comparisons are made to absorption imaging and a near-ideal detector using a signal-to-noise ratio that includes both statistical and structural noise. Similar to previous studies, heavy absorption filtration was needed to narrow the spectra at the expense of a large reduction in x-ray flux. Therefore, potential improvements using a chromatic multi-prism x-ray lens (MPL) for filtering were evaluated theoretically. The MPL offers a narrow tunable spectrum, and we show that the image quality can be improved compared to conventional filtering methods.
  •  
9.
  • Fredenberg, Erik, PhD, 1979-, et al. (författare)
  • Contrast-enhanced spectral mammography with a photon-counting detector
  • 2010
  • Ingår i: Medical physics (Lancaster). - : Wiley. - 0094-2405. ; 37:5, s. 2017-2029
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: Spectral imaging is a method in medical x-ray imaging to extract information about the object constituents by the material-specific energy dependence of x-ray attenuation. In particular, the detectability of a contrast agent can be improved over a lumpy background. We have investigated a photon-counting spectral imaging system with two energy bins for contrast-enhanced mammography. System optimization and the potential benefit compared to conventional non-energy-resolved imaging was studied.Methods: A framework for system characterization was set up that included quantum and anatomical noise, and a theoretical model of the system was benchmarked to phantom measurements.Results: It was found that optimal combination of the energy-resolved images corresponded approximately to minimization of the anatomical noise, and an ideal-observer detectability index could be improved more than a factor of two compared to absorption imaging in the phantom study. In the clinical case, an improvement close to 80% was predicted for an average glandularity breast, and a factor of eight for dense breast tissue. Another 70% was found to be within reach for an optimized system.Conclusions: Contrast-enhanced spectral mammography is feasible and beneficial with the current system, and there is room for additional improvements.
  •  
10.
  • Fredenberg, Erik, PhD, 1979-, et al. (författare)
  • Energy filtering with x-ray lenses: Optimization for photon-counting mammography
  • 2010
  • Ingår i: Radiation Protection Dosimetry. - : Oxford University Press (OUP). - 0144-8420 .- 1742-3406. ; 139, s. 339-342
  • Tidskriftsartikel (refereegranskat)abstract
    • Chromatic properties of the multi-prism and prism-array x-ray lenses (MPL and PAL) can potentially be utilized for efficient energy filtering and dose reduction in mammography. The line-shaped foci of the lenses are optimal for coupling to photon-counting silicon strip detectors in a scanning system. A theoretical model was developed and used to investigate the benefit of two lenses compared to an absorption-filtered reference system. The dose reduction of the MPL filter was 15% compared to the reference system at matching scan time, and the spatial resolution was higher. The dose of the PAL-filtered system was found to be 20% lower than for the reference system at equal scan time and resolution, and only 20% higher than for a monochromatic beam. An investigation of some practical issues remains, including the feasibility of brilliant-enough x-ray sources and manufacturing of a polymer PAL.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-10 av 29

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy