SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "L773:1434 6621 OR L773:1437 4331 "

Sökning: L773:1434 6621 OR L773:1437 4331

  • Resultat 1-25 av 201
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Aardal, Elisabeth, et al. (författare)
  • Cortisol in Saliva : Reference Ranges and Relation to Cortisol in Serum
  • 1995
  • Ingår i: European Journal of Clinical Chemistry and Clinical Biochemistry. - : Walter de Gruyter GmbH. - 0939-4974. ; 33, s. 927-932
  • Tidskriftsartikel (refereegranskat)abstract
    • The aim of this study was to establish morning and evening reference ranges for cortisol in saliva. Another objective was to compare the concentrations of the mainly free cortisol in saliva to those of total cortisol in serum as determined with a commercial radioimmunoassay. The concentrations were determined in matched samples of saliva and serum collected at 8am and 10pm from 197 healthy volunteers. The saliva samples were stable for at least 7 days at room temperature and for 9 months at —20 °C. Reference ranges, the central 95%, were estimated to 3.5—27.0 nmol/1 at 8 am and < 6.0 nmol/1 at 10 pm. The intra-assay coefficient of variation (CV) was below 5% and total CV below 10%. The relation between the cortisol concentrations in serum and saliva was nonlinear with r = 0.86 for serum concentrations < 450 nmol/1 and r = 0.44 for serum concentrations ^ 450 nmol/1. In conclusion, the satisfactory precision of the analysis and the simple non-invasive sampling procedure suggest that saliva may be used for cortisol measurements in situations where blood sampling is difficult to perform.
  •  
2.
  • Aardal-Eriksson, Elisabeth, et al. (författare)
  • Salivary cortisol : an alternative to serum cortisol determinations in dynamic function tests
  • 1998
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 36:4, s. 215-222
  • Tidskriftsartikel (refereegranskat)abstract
    • Salivary cortisol was measured as an alternative to serum cortisol as a marker for adrenocortical function following insulin tolerance test, corticotropin-releasing-hormone stimulation and adreno-corticotrophic hormone stimulation. During insulin tolerance test and corticotropin-releasing-hormone stimulation adreno-corticotrophic hormone was also measured. The tests were performed on healthy control subjects as well as on patients under investigation for various disturbances in the hypothalamic-pituitary-adrenocortical axis (insulin tolerance test: 3 controls on two occasions and 14 patients; corticotropin-releasing-hormone stimulation: 4 controls and 18 patients; adreno-corticotrophic hormone stimulation: 6 controls and 10 patients). Five patients underwent both insulin tolerance test and corticotropin-releasing-hormone stimulation. Using criteria for adequate cortisol response in serum, the patients were classified as good or poor responders. In 42 of the 45 tests performed the same conclusion as to cortisol status was drawn when based on serum and salivary cortisol responses. In healthy subjects and good responders the mean cortisol relative increase was greater in saliva than in serum in all three tests (p < 0.05). Characteristic of the results for the insulin tolerance test was a significant initial mean decrease (p < 0.05), not found in serum, and the highest observed salivary cortisol value was delayed for at least 30 minutes compared to that in serum. Plasma adreno-corticotrophic hormone correlated significantly with the cortisol concentrations determined 15 minutes later in serum (r = 0.54–0.64) and in saliva (r = 0.76–0.85). The more pronounced cortisol response in saliva than in serum and its closer correlation with adreno-corticotrophic hormone offer advantages over serum cortisol, suggesting salivary cortisol measurement may be used as an alternative parameter in dynamic endocrine tets.
  •  
3.
  • Andrade, Luis E. C., et al. (författare)
  • International consensus on antinuclear antibody patterns : definition of the AC-29 pattern associated with antibodies to DNA topoisomerase I
  • 2018
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : WALTER DE GRUYTER GMBH. - 1434-6621 .- 1437-4331. ; 56:10, s. 1783-1788
  • Tidskriftsartikel (refereegranskat)abstract
    • The indirect immunofluorescence assay (IFA) on HEp-2 cells is the reference method for autoantibody screening. The HEp-2 IFA pattern provides useful information on the possible autoantibodies in the sample. The International Consensus on Antinuclear Antibody Patterns (ICAP) initiative seeks to define and harmonize the nomenclature of HEp-2 IFA patterns. The most relevant and usual patterns have been assigned an alphanumeric code from anti-cell (AC)-1 to AC-28 and were organized into a classification algorithm (www.ANApatterns.org). The systemic sclerosis-associated autoantibodies to DNA topoisomerase I (Topo I) produce a peculiar composite 5-element HEp-2 IFA pattern (Topo I-like pattern) comprising the staining of the nucleus, metaphase chromatin plate, nucleolar organizing region, cytoplasm and nucleolus. In a recent assessment of the European Consensus Finding Study Group on autoantibodies, a well-defined anti-Topo I sample was blindly analyzed and classified according to ICAP AC patterns by 43 participant laboratories across Europe. There were wide variations among these laboratories in reporting nuclear, nucleolar and cytoplasmic patterns, indicating the inadequacy of the existing AC patterns to report the Topo I-like pattern. Several ICAP member laboratories independently demonstrated the overall consistency of the HEp-2 IFA Topo I-like pattern using HEp-2 slides from different manufacturers. The ICAP committee reviewed 24 candidate images and selected the four most representative images to be available on the ICAP website. The proper recognition of the AC-29 pattern should trigger suspicion of the presence of anti-Topo I antibodies, which may engender appropriate analyte-specific reflex tests to confirm the autoantibody specificity.
  •  
4.
  • Andreasson, Ulf, 1968, et al. (författare)
  • Assessing the commutability of candidate reference materials for the harmonization of neurofilament light measurements in blood
  • 2023
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter GmbH. - 1434-6621 .- 1437-4331. ; 61:7, s. 1245-1254
  • Tidskriftsartikel (refereegranskat)abstract
    • Objectives Neurofilament light chain (NfL) concentration in blood is a biomarker of neuro-axonal injury in the nervous system and there now exist several assays with high enough sensitivity to measure NfL in serum and plasma. There is a need for harmonization with the goal of creating a certified reference material (CRM) for NfL and an early step in such an effort is to determine the best matrix for the CRM. This is done in a commutability study and here the results of the first one for NfL in blood is presented.Methods Forty paired individual serum and plasma samples were analyzed for NfL on four different analytical platforms. Neat and differently spiked serum and plasma were evaluated for their suitability as a CRM using the difference in bias approach.Results The correlation between the different platforms with regards to measured NfL concentrations were very high (Spearman's rho >= 0.96). Samples spiked with cerebrospinal fluid (CSF) showed higher commutability compared to samples spiked with recombinant human NfL protein and serum seems to be a better choice than plasma as the matrix for a CRM.Conclusions The results from this first commutability study on NfL in serum/plasma showed that it is feasible to create a CRM for NfL in blood and that spiking should be done using CSF rather than with recombinant human NfL protein.
  •  
5.
  • Andreasson, Ulf, 1968, et al. (författare)
  • Commutability of the certified reference materials for the standardization of beta-amyloid 1-42 assay in human cerebrospinal fluid: lessons for tau and beta-amyloid 1-40 measurements
  • 2018
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter GmbH. - 1434-6621 .- 1437-4331. ; 56:12, s. 2058-2066
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: The core Alzheimer's disease cerebrospinal fluid (CSF) biomarkers total tau (T-tau), phosphorylated tau (P-tau), beta-amyloid 1-42 (A beta 42) and beta-amyloid 1-40 (A beta 40) are increasing in importance and are now part of the research criteria for the diagnosis of the disease. The main aim of this study is to evaluate whether a set of certified reference materials (CRMs) are commutable for A beta 42 and to serve as a feasibility study for the other markers. This property is a prerequisite for the establishment of CRMs which will then be used by manufacturers to calibrate their assays against. Once the preanalytical factors have been standardized and proper selection criteria are available for subject cohorts this harmonization between methods will allow for universal cut-offs to be determined. Methods: Thirty-four individual CSF samples and three different CRMs where analyzed for T-tau, P-tau, A beta 42 and A beta 40, using up to seven different commercially available methods. For A beta 40 and A beta 42 a mass spectrometry-based procedure was also employed. Results: There were strong pairwise correlations between the different methods (Spearman's p>0.92) for all investigated analytes and the CRMs were not distinguishable from the individual samples. Conclusions: This study shows that the CRMs are commutable for the different assays for A beta 42. For the other analytes the results show that it would be feasible to also produce CRMs for these. However, additional studies are needed as the concentration interval for the CRMs were selected based on A beta 42 concentrations only and did in general not cover satisfactory large concentration intervals for the other analytes.
  •  
6.
  • Apweiler, Rolf, et al. (författare)
  • Approaching clinical proteomics : current state and future fields of application in fluid proteomics
  • 2009
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 47:6, s. 724-744
  • Forskningsöversikt (refereegranskat)abstract
    • The field of clinical proteomics offers opportunities to identify new disease biomarkers in body fluids, cells and tissues. These biomarkers can be used in clinical applications for diagnosis, stratification of patients for specific treatment, or therapy monitoring. New protein array formats and improved spectrometry technologies have brought these analyses to a level with potential for use in clinical diagnostics. The nature of the human body fluid proteome with its large dynamic range of protein concentrations presents problems with quantitation. The extreme complexity of the proteome in body fluids presents enormous challenges and requires the establishment of standard operating procedures for handling of specimens, increasing sensitivity for detection and bioinformatical tools for distribution of proteomic data into the public domain. From studies of in vitro diagnostics, especially in clinical chemistry, it is evident that most errors occur in the preanalytical phase and during implementation of the diagnostic strategy. This is also true for clinical proteomics, and especially for fluid proteomics because of the multiple pretreatment processes. These processes include depletion of high-abundance proteins from plasma or enrichment processes for urine where biological variation or differences in proteolytic activities in the sample along with preanalytical variables such as inter- and intra-assay variability will likely influence the results of proteomics studies. However, before proteomic analysis can be introduced at a broader level into the clinical setting, standardization of the preanalytical phase including patient preparation, sample collection, sample preparation, sample storage, measurement and data analysis needs to be improved. In this review, we discuss the recent technological advances and applications that fulfil the criteria for clinical proteomics, with the focus on fluid proteomics. These advances relate to preanalytical factors, analytical standardization and quality-control measures required for effective implementation into routine laboratory testing in order to generate clinically useful information. With new disease biomarker candidates, it will be crucial to design and perform clinical studies that can identify novel diagnostic strategies based on these techniques, and to validate their impact on clinical decision-making.
  •  
7.
  • Arfvidsson, Berndt, et al. (författare)
  • S100B concentrations increase perioperatively in jugular vein blood despite limited metabolic and inflammatory response to clinically uneventful carotid endarterectomy
  • 2015
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter GmbH. - 1434-6621 .- 1437-4331. ; 53:1, s. 111-117
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: Our aim was to test the hypothesis that metabolic and inflammatory responses of the brain perioperatively during carotid endarterectomy (CEA) might affect blood brain barrier (BBB) integrity.Methods: Twenty patients with >70% stenosis of internal carotid artery (ICA) were prospectively included. Surgery was performed under general anaesthesia. Blood was sampled from ipsilateral internal jugular vein and radial artery: just before, during, and after ICA clamping S100B protein, glucose, lactate, 20 amino acids, and key cytokines were analysed.Results: Jugular vein S100B increased during clamping and reperfusion, while a marginal systemic increase was recorded, unrelated to stump pressure during clamping. Glucose increased during clamping in jugular vein blood and even more systemically, while jugular lactate values were higher than systemic values initially. Most amino acids did not differ significantly between jugular vein and systemic levels: glutamic acid and aspartic acid decreased during surgery while asparagine increased. Jugular vein interleukin (IL)-6 showed a transient non-significant increase during clamping and decreased systemically. IL-8 and IL-10 increased over time.Conclusions: Rising jugular vein S100B concentrations indicated reduced BBB integrity, and marginal secondary increase of S100B systemically. Limited ischaemic effects on the brain during cross-clamping, unrelated to S100B concentrations, were confirmed by lower brain glucose levels and higher lactate levels than in systemic blood. The lack of increased jugular vein glutamic acid disproves any major ischaemic brain injury following CEA. The inflammatory response was limited, did not differ greatly between jugular and systemic blood, and was unrelated to S100B.
  •  
8.
  • Arslan, Burak, 1990, et al. (författare)
  • Blood-based biomarkers in Alzheimer's disease - moving towards a new era of diagnostics
  • 2024
  • Ingår i: CLINICAL CHEMISTRY AND LABORATORY MEDICINE. - 1434-6621 .- 1437-4331.
  • Tidskriftsartikel (refereegranskat)abstract
    • Alzheimer's disease (AD), a primary cause of dementia globally, is traditionally diagnosed via cerebrospinal fluid (CSF) measures and positron emission tomography (PET). The invasiveness, cost, and limited accessibility of these methods have led to exploring blood-based biomarkers as a promising alternative for AD diagnosis and monitoring. Recent advancements in sensitive immunoassays have identified potential blood-based biomarkers, such as A beta 42/A beta 40 ratios and phosphorylated tau (p-tau) species. This paper briefly evaluates the clinical utility and reliability of these biomarkers across various AD stages, highlighting challenges like refining plasma A beta 42/A beta 40 assays and enhancing the precision of p-tau, particularly p-tau181, p-tau217, and p-tau231. The discussion also covers other plasma biomarkers like neurofilament light (NfL), glial fibrillary acidic protein (GFAP), and synaptic biomarkers, assessing their significance in AD diagnostics. The need for ongoing research and development of robust assays to match the performance of CSF and PET biomarkers is underscored. In summary, blood-based biomarkers are increasingly crucial in AD diagnosis, follow-up, prognostication, treatment response evaluation, and population screening, particularly in primary care settings. These developments are set to revolutionize AD diagnostics, offering earlier and more accessible detection and management options
  •  
9.
  • Arslan, Burak, 1990, et al. (författare)
  • Neurofilament light chain as neuronal injury marker - what is needed to facilitate implementation in clinical laboratory practice?
  • 2023
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter GmbH. - 1434-6621 .- 1437-4331. ; 61:7, s. 1140-1149
  • Tidskriftsartikel (refereegranskat)abstract
    • Neurobiomarkers have attracted significant attention over the last ten years. One promising biomarker is the neurofilament light chain protein (NfL). Since the introduction of ultrasensitive assays, NfL has been developed into a widely used axonal damage marker of relevance to the diagnosis, prognostication, follow-up, and treatment monitoring of a range of neurological disorders, including multiple sclerosis, amyotrophic lateral sclerosis, and Alzheimer's disease. The marker is increasingly used clinically, as well as in clinical trials. Even if we have validated precise, sensitive, and specific assays for NfL quantification in both cerebrospinal fluid and blood, there are analytical, as well as pre- and post-analytical aspects of the total NfL testing process, including biomarker interpretation, to consider. Although the biomarker is already in use in specialised clinical laboratory settings, a more general use requires some further work. In this review, we provide brief basic information and opinions on NfL as a biomarker of axonal injury in neurological diseases and pinpoint additional work needed to facilitate biomarker implementation in clinical practice.
  •  
10.
  • Arsov, Stefan, et al. (författare)
  • Advanced glycation end-products and skin autofluorescence in end-stage renal disease : a review
  • 2014
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter. - 1434-6621 .- 1437-4331. ; 52:1, SI, s. 11-20
  • Forskningsöversikt (refereegranskat)abstract
    • Chronic kidney disease (CKD), especially in its end stage, is marked by extremely high cardiovascular rates of morbidity and mortality; hemodialysis patients have a five-fold shorter life expectancy than healthy subjects of the same age. In CKD the metabolic products that accumulate in the body are so-called uremic toxins. These include advanced glycation end-products (AGE). AGE levels are markedly increased in CKD patients not only because of impaired excretion but also because of increased production. AGE formation has initially been described as a non-enzymatic reaction between proteins and glucose in the so-called Maillard reaction, but they are also more rapidly formed during oxidative stress and subsequent formation of reactive carbonyl compounds like (methyl) glyoxal. AGE accumulate in tissue where they cross-link with proteins, e. g., collagen, inducing tissue stiffening of blood vessels and skin. They may also interact with receptor of AGE (RAGE) and other receptors, which lead to activation of intracellular transduction mechanisms resulting in cytokine release and further tissue damage in CKD. The accumulation of AGE in the skin can be measured non-invasively using autofluorescence. The skin autofluorescence is a strong marker of cardiovascular mortality in CKD. The focus of this review is on the role of tissue and plasma AGE, and of skin autofluorescence as a proxy of tissue AGE accumulation, in the increase in cardiovascular disease in end stage renal disease (ESRD). This review will also present the possibility of reducing the AGE accumulation in ESRD patients using the following five methods: 1) use of low AGE peritoneal dialysis solutions; 2) use of advanced hemodialysis techniques; 3) use of AGE reducing drugs; 4) optimizing the nutrition of hemodialysis patients; and 5) renal transplantation.
  •  
11.
  •  
12.
  • Ben, Rayana M.C., et al. (författare)
  • Guidelines for sampling, measuring and reporting ionized magnesium in undiluted serum, plasma or blood : International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)
  • 2005
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 43:5, s. 564-569
  • Tidskriftsartikel (refereegranskat)abstract
    • All analyzers with ion-selective electrodes for ionized magnesium (iMg) should yield comparable and unbiased results. The prerequisite to achieve this goal is to reach consensus on sampling, measurement and reporting. The recommended guidelines for sampling, measurement and reporting iMg in plasma ("plasma" refers to circulating plasma and the forms in which it is sampled: the plasma phase of anticoagulated whole blood, plasma separated from blood cells, or serum) or blood, referring to the substance concentration of iMg in the calibrants, will provide results for iMg that are approximately 3% greater than its true concentration, and 4% less than its true molality. Binding of magnesium to proteins and ligands in plasma and blood is pH-dependent. Therefore, pH should be simultaneously measured to allow adjustment of iMg concentration to pH 7.4. The substance concentration of iMg may be physiologically and consequently clinically more relevant than the substance concentration of total magnesium. © 2005 by Walter de Gruyter.
  •  
13.
  • Ben Rayana, Mohammed C., et al. (författare)
  • IFCC guideline for sampling, measuring and reporting ionized magnesium in plasma
  • 2008
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 46:1, s. 21-26
  • Tidskriftsartikel (refereegranskat)abstract
    • Analyzers with ion-selective electrodes (ISEs) for ionized magnesium (iMg) should yield comparable and unbiased results for iMg. This IFCC guideline on sampling, measuring and reporting iMg in plasma provides a prerequisite to achieve this goal [in this document, "plasma" refers to circulating plasma and the forms in which it is sampled, namely the plasma phase of anticoagulated whole blood (or "blood"), plasma separated from blood cells, or serum]. The guideline recommends measuring and reporting ionized magnesium as a substance concentration relative to the substance concentration of magnesium in primary aqueous calibrants with magnesium, sodium, and calcium chloride of physiological ionic strength. The recommended name is "the concentration of ionized magnesium in plasma". Based on this guideline, results will be approximately 3% higher than the true substance concentration and 4% lower than the true molality in plasma. Calcium ions interfere with all current magnesium ion-selective electrodes (Mg-ISEs), and thus it is necessary to determine both ions simultaneously in each sample and correct the result for Ca2+ interference. Binding of Mg in plasma is pH-dependent. Therefore, pH should be measured simultaneously with iMg to allow adjustment of the result to pH 7.4. The concentration of iMg in plasma may be physiologically and clinically more relevant than the concentration of total magnesium. Furthermore, blood-gas analyzers or instruments for point-of-care testing are able to measure plasma iMg using whole blood (with intact blood cells) as the sample, minimizing turnaround time compared to serum and plasma, which require removal of blood cells.
  •  
14.
  • Ben Rayana, Mohammed C, et al. (författare)
  • Recommendation for measuring and reporting chloride by ISEs in undiluted serum, plasma or blood
  • 2006
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 44:3, s. 346-352
  • Tidskriftsartikel (refereegranskat)abstract
    • The proposed recommendation for measuring and reporting chloride in undiluted plasma† or blood by ion-selective electrodes (ISEs) will provide results that are identical to chloride concentrations measured by coulometry for standardized normal plasma or blood samples. It is applicable to all current ISEs dedicated to chloride measurement in undiluted samples that meet the requirements. However, in samples with reduced water concentration, results by coulometry are lower than by ion-selective electrode due to volume displacement. The quantity measured by this standardized ISE procedure is called the ionized chloride concentration. It may be clinically more relevant than the chloride concentration as determined by coulometry, photometry or by ISE after dilution of the sample. © 2006 by Walter de Gruyter.
  •  
15.
  • Bjerke, Maria, 1977, et al. (författare)
  • Assessing the commutability of reference material formats for the harmonization of amyloid beta measurements.
  • 2016
  • Ingår i: Clinical chemistry and laboratory medicine : CCLM / FESCC. - : Walter de Gruyter GmbH. - 1437-4331 .- 1434-6621. ; 54:7, s. 1177-91
  • Tidskriftsartikel (refereegranskat)abstract
    • BACKGROUND: The cerebrospinal fluid (CSF) amyloid-β (Aβ42) peptide is an important biomarker for Alzheimer's disease (AD). Variability in measured Aβ42 concentrations at different laboratories may be overcome by standardization and establishing traceability to a reference system. Candidate certified reference materials (CRMs) are validated herein for this purpose. METHODS: Commutability of 16 candidate CRM formats was assessed across five CSF Aβ42 immunoassays and one mass spectrometry (MS) method in a set of 48 individual clinical CSF samples. Promising candidate CRM formats (neat CSF and CSF spiked with Aβ42) were identified and subjected to validation across eight (Elecsys, EUROIMMUN, IBL, INNO-BIA AlzBio3, INNOTEST, MSD, Simoa, and Saladax) immunoassays and the MS method in 32 individual CSF samples. Commutability was evaluated by Passing-Bablok regression and the candidate CRM termed commutable when found within the prediction interval (PI). The relative distance to the regression line was assessed. RESULTS: The neat CSF candidate CRM format was commutable for almost all method comparisons, except for the Simoa/MSD, Simoa/MS and MS/IBL where it was found just outside the 95% PI. However, the neat CSF was found within 5% relative distance to the regression line for MS/IBL, between 5% and 10% for Simoa/MS and between 10% and 15% for Simoa/MSD comparisons. CONCLUSIONS: The neat CSF candidate CRM format was commutable for 33 of 36 method comparisons, only one comparison more than expected given the 95% PI acceptance limit. We conclude that the neat CSF candidate CRM can be used for value assignment of the kit calibrators for the different Aβ42 methods.
  •  
16.
  • Björk, Jonas, et al. (författare)
  • Accuracy of GFR estimating equations combining standardized cystatin C and creatinine assays: a cross-sectional study in Sweden
  • 2015
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter GmbH. - 1434-6621 .- 1437-4331. ; 53:3, s. 403-414
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: The recently established international cystatin C calibrator makes it possible to develop non-laboratory specific glomerular filtration rate (GFR) estimating (eGFR) equations. This study compares the performance of the arithmetic mean of the revised Lund-Malmo creatinine and CAPA cystatin C equations (MEAN(LM-REV+CAPA)), the arithmetic mean of the Chronic Kidney Disease Epidemiology Collaboration equation (CKD-EPI) creatinine and cystatin C equations (MEAN(CKD-EPI)), and the composite CKD-EPI equation (CKD-EPICREA+CYSC) with the corresponding single marker equations using internationally standardized calibrators for both cystatin C and creatinine. Methods: The study included 1200 examinations in 1112 adult Swedish patients referred for measurement of GFR (mGFR) 2008-2010 by plasma clearance of iohexol (median 51 mL/min/1.73 m(2)). Bias, precision (interquartile range, IQR) and accuracy (percentage of estimates +/- 30% of mGFR; P-30) were compared. Results: Combined marker equations were unbiased and had higher precision and accuracy than single marker equations. Overall results of MEAN(LM-REV+CAPA)/MEAN(CKD-EPI)/CKD-EPICREA+CYSC were: median bias -2.2%/-0.5%/-1.6%, IQR 9.2/9.2/8.8 mL/min/1.73 m(2), and P-30 91.3%/91.0%/91.1%. The P-30 figures were about 7-14 -percentage points higher than the single marker equations. The combined equations also had a more stable performance across mGFR, age and BMI intervals, generally with P-30 >= 90% and never <80%. Combined equations reached P-30 of 95% when the difference between eGFR(CREA) and eGFR(CYSC) was <10% but decreased to 82% at a difference of >= 40%. Conclusions: Combining cystatin C and creatinine assays improves GFR estimations with P-30 >= 90% in adults. Reporting estimates of both single and combined marker equations in clinical settings makes it possible to assess the validity of the combined equation based on the agreement between the single marker equations.
  •  
17.
  • Björk, Jonas, et al. (författare)
  • GFR estimation based on standardized creatinine and cystatin C : A European multicenter analysis in older adults
  • 2018
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter GmbH. - 1437-4331 .- 1434-6621. ; 56:3, s. 422-435
  • Tidskriftsartikel (refereegranskat)abstract
    • Although recommended by the Kidney Disease Improving Global Outcomes, the Chronic Kidney Disease Epidemiology Collaboration (CKD-EPICR) creatinine equation was not targeted to estimate glomerular filtration rate (eGFR) among older adults. The Berlin Initiative Study (BIS1CR) equation was specifically developed in older adults, and the Lund-Malmö revised (LMRCR) and the Full Age Spectrum (FASCR) equations have shown promising results in older adults. Our aim was to validate these four creatinine equations, including addition of cystatin C in a large multicenter cohort of Europeans ≥70 years. A total of 3226 individuals (2638 with cystatin C) underwent GFR measurement (mGFR; median, 44 mL/min/1.73 m2) using plasma iohexol clearance. Bias, precision (interquartile range [IQR]), accuracy (percent of estimates ±30% of mGFR, P30), eGFR accuracy diagrams and probability diagrams to classify mGFR<45 mL/min/1.73 m2 were compared. The overall results of BIS1CR/CKD-EPICR/FASCR/LMRCR were as follows: median bias, 1.7/3.6/0.6/-0.7 mL/min/1.73 m2; IQR, 11.6/12.3/11.1/10.5 mL/min/1.73 m2; and P30, 77.5%/76.4%/80.9%/83.5% (significantly higher for LMR, p<0.001). Substandard P30 (<75%) was noted for all equations at mGFR<30 mL/min/1.73 m2, and at body mass index values <20 and ≥35 kg/m2. LMRCR had the most stable performance across mGFR subgroups. Only LMRCR and FASCR had a relatively constant small bias across eGFR levels. Probability diagrams exhibited wide eGFR intervals for all equations where mGFR<45 could not be confidently ruled in or out. Adding cystatin C improved P30 accuracy to 85.7/86.8/85.7/88.7 for BIS2CR+CYS/CKD-EPICR+CYS/FASCR+CYS/MEANLMR+CAPA. LMRCR and FASCR seem to be attractive alternatives to CKD-EPICR in estimating GFR by creatinine-based equations in older Europeans. Addition of cystatin C leads to important improvement in estimation performance.
  •  
18.
  • Blennow, Kaj, 1958, et al. (författare)
  • Second-generation Elecsys cerebrospinal fluid immunoassays aid diagnosis of early Alzheimer's disease.
  • 2023
  • Ingår i: Clinical chemistry and laboratory medicine. - : Walter de Gruyter GmbH. - 1437-4331 .- 1434-6621. ; 61:2, s. 234-244
  • Tidskriftsartikel (refereegranskat)abstract
    • Timely diagnosis of Alzheimer's disease (AD) is critical for appropriate treatment/patient management. Cerebrospinal fluid (CSF) biomarker analysis is often used to aid diagnosis. We assessed analytical performance of second-generation (Gen II) Elecsys® CSF immunoassays (Roche Diagnostics International Ltd), and adjusted existing cut-offs, to evaluate their potential utility in clinical routine.Analytical performance was assessed using CSF samples measured with Elecsys CSF Gen II immunoassays on cobas e analyzers. Aβ42 Gen I/Gen II immunoassay method comparisons were performed (Passing-Bablok regression). Cut-off values were adjusted using estimated bias in biomarker levels between BioFINDER protocol aliquots/Gen I immunoassays and Gen II protocol aliquots/immunoassays. Distribution of Gen II immunoassay values was evaluated in AD, mild cognitive impairment (MCI), and cognitively normal cohorts; percentage observations outside the measuring range were derived.The Gen II immunoassays demonstrated good analytical performance, including repeatability, intermediate precision, lot-to-lot agreement (Pearson's r:≥0.999), and platform agreement (Pearson's r:≥0.995). Aβ42 Gen I/Gen II immunoassay measurements were strongly correlated (Pearson's r: 0.985-0.999). Aβ42 Gen II immunoassay cut-offs were adjusted to 1,030 and 800ng/L, and pTau181/Aβ42 ratio cut-offs to 0.023 and 0.029, for Gen II and I protocols, respectively. No observations were below the lower limit of the measuring range; above the upper limit, there were none from the AD cohort, and 2.6 and 6.8% from the MCI and cognitively normal cohorts, respectively.Our findings suggest that the Gen II immunoassays have potential utility in clinical routine to aid diagnosis of AD.
  •  
19.
  • Bonroy, Carolien, et al. (författare)
  • Detection of antinuclear antibodies : recommendations from EFLM, EASI and ICAP
  • 2023
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter. - 1434-6621 .- 1437-4331. ; 61:7, s. 1167-1198
  • Tidskriftsartikel (refereegranskat)abstract
    • Objectives: Antinuclear antibodies (ANA) are important for the diagnosis of various autoimmune diseases. ANA are usually detected by indirect immunofluorescence assay (IFA) using HEp-2 cells (HEp-2 IFA). There are many variables influencing HEp-2 IFA results, such as subjective visual reading, serum screening dilution, substrate manufacturing, microscope components and conjugate. Newer developments on ANA testing that offer novel features adopted by some clinical laboratories include automated computer-assisted diagnosis (CAD) systems and solid phase assays (SPA).Methods: A group of experts reviewed current literature and established recommendations on methodological aspects of ANA testing. This process was supported by a two round Delphi exercise. International expert groups that participated in this initiative included (i) the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group “Autoimmunity Testing”; (ii) the European Autoimmune Standardization Initiative (EASI); and (iii) the International Consensus on ANA Patterns (ICAP).Results: In total, 35 recommendations/statements related to (i) ANA testing and reporting by HEp-2 IFA; (ii) HEp-2 IFA methodological aspects including substrate/conjugate selection and the application of CAD systems; (iii) quality assurance; (iv) HEp-2 IFA validation/verification approaches and (v) SPA were formulated. Globally, 95% of all submitted scores in the final Delphi round were above 6 (moderately agree, agree or strongly agree) and 85% above 7 (agree and strongly agree), indicating strong international support for the proposed recommendations.Conclusions: These recommendations are an important step to achieve high quality ANA testing.
  •  
20.
  •  
21.
  • Bäcklund, Nils, et al. (författare)
  • Salivary cortisol and cortisone in diagnosis of Cushing's syndrome - a comparison of six different analytical methods
  • 2023
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter. - 1434-6621 .- 1437-4331. ; 61:10, s. 1780-1791
  • Tidskriftsartikel (refereegranskat)abstract
    • Objectives: Salivary cortisol and cortisone at late night and after dexamethasone suppression test (DST) are increasingly used for screening of Cushing's syndrome (CS). We aimed to establish reference intervals for salivary cortisol and cortisone with three liquid chromatography-tandem mass spectrometry (LC-MS/MS) techniques and for salivary cortisol with three immunoassays (IAs), and evaluate their diagnostic accuracy for CS.Methods: Salivary samples at 08:00 h, 23:00 h and 08:00 h after a 1-mg DST were collected from a reference population (n=155) and patients with CS (n=22). Sample aliquots were analyzed by three LC-MS/MS and three IA methods. After establishing reference intervals, the upper reference limit (URL) for each method was used to calculate sensitivity and specificity for CS. Diagnostic accuracy was evaluated by comparing ROC curves.Results: URLs for salivary cortisol at 23:00 h were similar for the LC-MS/MS methods (3.4-3.9 nmol/L), but varied between IAs: Roche (5.8 nmol/L), Salimetrics (4.3 nmol/L), Cisbio (21.6 nmol/L). Corresponding URLs after DST were 0.7-1.0, and 2.4, 4.0 and 5.4 nmol/L, respectively. Salivary cortisone URLs were 13.5-16.6 nmol/L at 23:00 h and 3.0-3.5 nmol/L at 08:00 h after DST. All methods had ROC AUCs =0.96.Conclusions: We present robust reference intervals for salivary cortisol and cortisone at 08:00 h, 23:00 h and 08:00 h after DST for several clinically used methods. The similarities between LC-MS/MS methods allows for direct comparison of absolute values. Diagnostic accuracy for CS was high for all salivary cortisol and cortisone LC-MS/MS methods and salivary cortisol IAs evaluated.
  •  
22.
  • Bölenius, Karin, et al. (författare)
  • Minor improvement of venous blood specimen collection practices in primary health care after a large-scale educational intervention
  • 2013
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter GmbH. - 1434-6621 .- 1437-4331. ; 51:2, s. 303-310
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: Venous blood specimen collection is a common health care practice that has to follow strict guidelines, non-compliance among sampling staff may compromise patient safety. We evaluated a large-scale 2 h educational intervention that emphasised guideline adherence to assess possible improvements of venous blood specimen collection practices.Methods: Blood specimen haemolysis is usually caused by inadequate venous blood specimen collection and handling, reflecting overall pre-analytical handling. We monitored haemolysis of serum samples with haemolysis index corresponding to ≥150 mg/L of free haemoglobin for specimens sent from 11 primary health care centres and analysed on a Vitros 5,1 clinical chemistry analyser before (2008, n=6652 samples) and after (2010, n=6121 samples) the intervention.Results: The total percentage of haemolysed specimens was 11.8% compared to 10.5% (p=0.022) before the intervention. As groups, rural primary health care centres demonstrated a significant reduction [Odds ratios (OR)=0.744] of haemolysed specimens after intervention, whereas urban primary health care centres demonstrated a significant increase (OR=1.451) of haemolysis.Conclusions: A large-scale 2 h educational intervention to make venous blood specimen collection staff comply with guideline practices had minor effects on collection practices. Educational interventions may be effective in wards/care centres demonstrating venous blood specimen collection practices with larger deviations from guidelines.
  •  
23.
  • Börjel, Anna K., et al. (författare)
  • Novel mutations in the 5'-UTR of the FOLR1 gene
  • 2006
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - 1434-6621 .- 1437-4331. ; 44:2, s. 161-167
  • Tidskriftsartikel (refereegranskat)abstract
    • We have previously reported two novel mutations in the 5'-untranslated region (UTR) of the gene for folate receptor-alpha (FOLR1). In our search for additional mutations, 92 patient samples with elevated levels of homocysteine were screened by single-strand conformation polymorphism (SSCP) between nt -425 and -782, and -712 and -1110. Between nt -425 and -782 we did not find any mutations. Between nt -712 and -1110 there were three novel mutations. One subject had two mutations very close to each other, c.-856C>T and c.-921T>C. Two subjects had a c.-1043G>A mutation. To get an idea of the prevalence of FOLR1 mutations in an unselected population, we also screened 692 healthy school children for mutations. In this cohort, between nt -188 and +272 we discovered one novel mutation, a single nucleotide substitution, c.-18C>T, in addition to five children with the 25-bp deletion mutation previously described by us. Thus, so far we have discovered six novel mutations in the 5'-UTR region of the gene for folate receptor-alpha. We genotyped all 17 subjects with a FOLR1 mutation for the methylenetetrahydrofolate reductase (MTHFR) 677C>T polymorphism, and developed new single-nucleotide polymorphism (SNP) genotyping protocols for MTHFR 1298A>C and 1793G>A utilising Pyrosequencing technology. None of the 17 subjects had the 677TT genotype, which ruled out this as a cause of elevated homocysteine levels, which was observed in some of the subjects. Further studies of mutations in the 5'-UTR of FOLR1, and in particular of their interplay with folate intake status, are warranted.
  •  
24.
  • Cadamuro, Janne, et al. (författare)
  • Preanalytical quality improvement - an interdisciplinary journey, on behalf of the European Federation for Clinical Chemistry and Laboratory Medicine (EFLM) Working Group for Preanalytical Phase (WG-PRE)
  • 2022
  • Ingår i: Clinical Chemistry and Laboratory Medicine. - : Walter de Gruyter. - 1434-6621 .- 1437-4331. ; 60:5, s. 662-668
  • Tidskriftsartikel (refereegranskat)abstract
    • Since the beginning of laboratory medicine, the main focus was to provide high quality analytics. Over time the importance of the extra-analytical phases and their contribution to the overall quality became evident. However, as the initial preanalytical processes take place outside of the laboratory and mostly without its supervision, all professions participating in these process steps, from test selection to sample collection and transport, need to engage accordingly. Focusing solely on intra-laboratory processes will not be sufficient to achieve the best possible preanalytical quality. The Working Group for the Preanalytical Phase (WG-PRE) of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) has provided several recommendations, opinion papers and scientific evidence over the past years, aiming to standardize the preanalytical phase across Europe. One of its strategies to reach this goal are educational efforts. As such, the WG-PRE has organized five conferences in the past decade with the sole focus on preanalytical quality. This year's conference mainly aims to depict the views of different professions on preanalytical processes in order to acquire common ground as basis for further improvements. This article summarizes the content of this 6th preanalytical conference.
  •  
25.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-25 av 201
Typ av publikation
tidskriftsartikel (188)
forskningsöversikt (8)
konferensbidrag (4)
recension (1)
Typ av innehåll
refereegranskat (183)
övrigt vetenskapligt/konstnärligt (18)
Författare/redaktör
Zetterberg, Henrik, ... (20)
Blennow, Kaj, 1958 (16)
Grankvist, Kjell (14)
Hultberg, Björn (12)
Kallner, A (12)
Larsson, Anders (10)
visa fler...
Theodorsson, Elvar (10)
Andreasson, Ulf, 196 ... (7)
Grubb, Anders (6)
Nilsson, Torbjörn K (6)
Church, Stephen (6)
Hultdin, Johan (6)
Hammarsten, Ola (5)
Damoiseaux, Jan (5)
Lindahl, Bertil, 195 ... (5)
Herold, Manfred (4)
Rönnelid, Johan (4)
Björk, Jonas (4)
Ridefelt, Peter (4)
Gustafson, Lars (4)
Nilsson, Karin (4)
Sandberg, Sverre (4)
Gellerstedt, Martin, ... (4)
Nyman, Ulf (4)
Plebani, Mario (4)
Aakre, K. M. (3)
Collinson, P. (3)
Gil, JP (3)
Ribeiro, V (3)
Yu, A (3)
Portelius, Erik, 197 ... (3)
Bossuyt, Xavier (3)
Fritzler, Marvin J. (3)
Chan, Edward K. L. (3)
Eggers, Kai M., 1962 ... (3)
Lindahl, Bertil (3)
Helander, A (3)
Jonasson, Torfi (3)
Öhlin, Hans (3)
Andrade, Luis E. C. (3)
Satoh, Minoru (3)
Arslan, Burak, 1990 (3)
Kuhle, J. (3)
Mattsson, Niklas, 19 ... (3)
Whitfield, John B. (3)
Helander, Anders (3)
Larsson, Lasse (3)
Eriksson, Catharina, ... (3)
Panteghini, Mauro (3)
Zegers, Ingrid (3)
visa färre...
Lärosäte
Karolinska Institutet (54)
Göteborgs universitet (39)
Lunds universitet (39)
Linköpings universitet (33)
Umeå universitet (32)
Uppsala universitet (32)
visa fler...
Örebro universitet (6)
Högskolan Väst (4)
Högskolan i Skövde (4)
Högskolan Kristianstad (1)
Malmö universitet (1)
visa färre...
Språk
Engelska (201)
Forskningsämne (UKÄ/SCB)
Medicin och hälsovetenskap (125)
Naturvetenskap (10)
Teknik (7)

År

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy