SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "WFRF:(Alm Carlsson G.) "

Sökning: WFRF:(Alm Carlsson G.)

  • Resultat 1-14 av 14
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Carlsson, C.A., et al. (författare)
  • An instrument for measuring ambient dose equivalent, H*(10)
  • 1996
  • Ingår i: Radiation Protection Dosimetry. - 0144-8420 .- 1742-3406. ; 67:1, s. 33-39
  • Tidskriftsartikel (refereegranskat)abstract
    • The design and calibration of a small and simple instrument for measuring the ambient dose equivalent, H*(10), in photon fields is described. Comprising a thermoluminescence LiF dosemeter inside a 20 mm diameter PMMA sphere, it is capable of measuring the ambient dose equivalent with a nearly isotropic response. In the interval 0.1-100 mSv and for the energy range 30 keV to 1.25 MeV the energy response is within -31% and +15% relative to that of 137Cs gamma radiation (662 keV). In practical use, it is therefore sufficient to calibrate the instrument in a 137Cs gamma field using the corresponding conversion coefficient H*(10)/Kair taken from tabulations. The possibility of using the instrument to monitor the ambient dose equivalent for energies above 1.25 MeV is discussed and indicates that the range of applicability can be extended to 4.4 MeV with an energy response within -10% relative to 662 keV.
  •  
2.
  • Nilsson, Håkan, et al. (författare)
  • Measurements of energy spectra from high dose rate 192Ir sources with a compton scattering spectrometer
  • 1994
  • Ingår i: Measurement assurance in dosimetry. - 9201001940 ; , s. 289-297
  • Konferensbidrag (övrigt vetenskapligt/konstnärligt)abstract
    • A Compton scattering spectrometer has been used for spectral measurement of high dose rate (HDR) 192Ir sources. On the basis of the Compton formula a measured distribution of scattered photons is used for the calculation of the primary spectrum leaving the source. The two main reasons for measuring the photon energy distribution from such HDR sources are, firstly, to obtain accurate input for Monte Carlo calculations of the dose distribution and, secondly, to calibrate ionization chambers. The lack of spectral information causes calibration laboratories great difficulties in such work. A third possible reason concerns quality assurance with respect to source impurities, etc. The measured spectrum shows good agreement with the spectrum expected from theoretical considerations.
  •  
3.
  •  
4.
  •  
5.
  •  
6.
  •  
7.
  • Hammersberg, P., et al. (författare)
  • A theoretical model for determination of the optimal irradiation conditions for computerised tomography
  • 1995
  • Ingår i: Insight. - 1354-2575 .- 1754-4904. ; 37:12, s. 978-985
  • Tidskriftsartikel (refereegranskat)abstract
    • Image quality in Computerised Tomography (CT) depends strongly on the quality of the CT-projection data. These depend on sample composition and geometry, contrasting details within the sample and the equipment used, i.e. X-ray spectra, filtration, detector response and geometry. This paper focuses on the problem of selecting the optimal physical parameters to maximise the signal-to-noise in CT projection data (SNRCT) between a contrasting detail and the surrounding sample for CT-scanners equipped with poly-energetic X-ray sources (conventional X-ray tubes) and energy-integrating detector systems (image intensifier and optical video chain). The work includes the derivation and verification ofa theoretical model for SNRCT which can be used for predicting the optimal physical parameters for specific imaging tasks. It is shown that simplified calculations valid for mono-energetic X-ray sources and/or photon counting detectors do not correctly predict the optimal settings. This study also includes measurements of the actual X-ray source and photon transport Monte Carlo simulations of the response of the detector system.
  •  
8.
  •  
9.
  •  
10.
  • Sandborg, Michael, 1961-, et al. (författare)
  • Demonstration of correlations between clinical and physical image quality measures in chest and lumbar spine screen-film radiography
  • 2001
  • Ingår i: British Journal of Radiology. - : British Institute of Radiology. - 0007-1285 .- 1748-880X. ; 74:882, s. 520-528
  • Tidskriftsartikel (refereegranskat)abstract
    • The ability to predict clinical image quality from physical measures is useful for optimization in diagnostic radiology. In this work, clinical and physical assessments of image quality are compared and correlations between the two are derived. Clinical assessment has been made by a group of expert radiologists who evaluated fulfilment of the European image criteria for chest and lumbar spine radiography using two scoring methods: image criteria score (ICS) and visual grading analysis score (VGAS). Physical image quality measures were calculated using a Monte Carlo simulation model of the complete imaging system. This model includes a voxelized male anatomy and was used to calculate contrast and signal-to-noise ratio of various important anatomical details and measures of dynamic range. Correlations between the physical image quality measures on the one hand and the ICS and VGAS on the other were sought. 16 chest and 4 lumbar spine imaging system configurations were compared in frontal projection. A statistically significant correlation with clinical image quality was found in chest posteroanterior radiography for the contrast of blood vessels in the retrocardiac area and a measure of useful dynamic range. In lumbar spine anteroposterior radiography, a similar significant correlation with clinical image quality was found between the contrast and signal-to-noise ratio of the trabecular structures in the L1-L5 vertebrae. The significant correlation shows that clinical image quality can, at least in some cases, be predicted from appropriate measures of physical image quality.
  •  
11.
  •  
12.
  •  
13.
  • Tingberg, Anders, et al. (författare)
  • Comparison of two methods for evaluation of image quality of lumbar spine radiographs
  • 2004
  • Ingår i: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. - : SPIE. - 1605-7422. ; 5372, s. 251-262
  • Tidskriftsartikel (övrigt vetenskapligt/konstnärligt)abstract
    • Purpose To evaluate the image quality of clinical radiographs with two different methods, and to find correlations between the two methods. Methods Based on fifteen lumbar spine radiographs, two new sets of images were created. A hybrid image set was created by adding two distributions of artificial lesions to each original image. The image quality parameters spatial resolution and noise were manipulated and a total of 210 hybrid images were created. A set of 105 disease-free images was created by applying the same combinations of spatial resolution and noise to the original images. The hybrid images were evaluated with the free-response forced error experiment (FFE) and the normal images with visual grading analysis (VGA) by nine experienced radiologists. Results The VGA study showed that images with low noise are preferred over images with higher noise levels. The alteration of the MTF had a limited influence on the VGA score. For the FFE study the visibility of the lesions was independent of the spatial resolution and the noise level. Discussion In this study we found no correlation between the two methods, probably because the detectability of the artificial lesions was not influenced by the manipulations of noise level and resolution. Hence, the detection of lesions in lumbar spine radiography may not be a quantum-noise limited task. The results show the strength of the VGA technique in terms of detecting small changes in the two image quality parameters. The method is more robust and has a higher statistical power than the ROC related method and could therefore, in some cases, be more suitable for use in optimization studies.
  •  
14.
  • Tingberg, Anders, et al. (författare)
  • Influence of the characteristic curve on the clinical image quality of lumbar spine and chest radiographs.
  • 2004
  • Ingår i: The British journal of radiology. - : British Institute of Radiology. - 0007-1285 .- 1748-880X. ; 77:915, s. 204-15
  • Tidskriftsartikel (refereegranskat)abstract
    • The "European Guidelines on Quality Criteria for Diagnostic Radiographic Images" do not address the choice of the film characteristic (H&D) curve, which is an important parameter for the description of a radiographic screen-film system. The image contrast of clinical lumbar spine and chest radiographs was altered by digital image processing techniques, simulating images with different H&D curves, both steeper and flatter than the original. The manipulated images were printed on film for evaluation. Seven experienced radiologists evaluated the clinical image quality by analysing the fulfilment of the European Image Criteria (ICS) and by visual grading analysis (VGA) of in total 224 lumbar spine and 360 chest images. A parallel study of the effect of the H&D curve has also been made using a theoretical model. The contrast (DeltaOD) of relevant anatomical details was calculated, using a Monte Carlo simulation-model of the complete imaging system including a 3D voxel phantom of a patient. Correlations between the calculated contrast and the radiologists' assessment by VGA were sought. The results of the radiologists' assessment show that the quality in selected regions of lumbar spine and chest images can be significantly improved by the use of films with a steeper H&D curve compared with the standard latitude film. Significant (p<0.05) correlations were found between the VGA results and the calculations of the contrast of transverse processes and trabecular details in the lumbar spine vertebrae, and with the contrast of blood vessels in the retrocardiac area of the chest.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-14 av 14

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy