1. |
- Bech, Martin, et al.
(författare)
-
Results from the first preclinical CT scanner with grating based phase contrast and a rotating gantry
- 2012
-
Ingår i: International Workshop on X-Ray and Neutron Phase Imaging with Gratings. - : AIP. - 1551-7616 .- 0094-243X. - 9780735410725 ; 1466, s. 130-136
-
Konferensbidrag (refereegranskat)abstract
- After successful demonstrations of soft-tissue phase-contrast imaging with grating interferometers at synchrotron radiation sources and at laboratory based x-ray tubes, a first preclinical CT scanner with grating based phase contrast imaging modality has been constructed. The rotating gantry is equipped with a three-grating interferometer, a 50 watt tungsten anode source and a Hamamatsu flat panel detector. The total length of the interferometer is 45 cm, and the bed of the scanner is optimized for mice, with a scanning diameter of 35 mm. From one single scan both phase-contrast and standard attenuation based tomography can be attained, providing an overall gain in image contrast.
|
|