1.
  • Flodin, Mats, et al. (author)
  • Evaluation of Gentian cystatin C reagent on Abbott Ci8200 and calculation of glomerular filtration rate expressed in mL/min/1.73 m(2) from the cystatin C values in mg/L
  • 2007
  • In: Scandinavian Journal of Clinical and Laboratory Investigation. - : Informa UK Limited. - 0036-5513 .- 1502-7686. ; 67:5, s. 560-567
  • Journal article (peer-reviewed)abstract
    • OBJECTIVE: Estimation of the glomerular filtration rate (GFR) is essential when evaluating patients with kidney disease and treating patients with drugs eliminated from the circulation by the kidneys. Cystatin C has been shown in several studies to be superior to creatinine in the estimation of GFR. At our hospitals, there is an increasing demand for cystatin C and at present we perform approximately 1500 cystatin C analyses a month. We thus need the assay available 24 h/day and to have it on our routine chemistry instrument to minimize handling time per test and time to reported test results. MATERIAL AND METHODS: We have evaluated a new cystatin C immunoassay from Gentian (Gentian, Moss, Norway) on Architect ci8200 (Abbott Laboratories, Abbott Park, Ill., USA). A prerequisite at our hospital is that cystatin C results are reported as a calculated GFR in mL/min/1.73 m(2), so we also made a comparison with iohexol clearance. RESULTS: The Gentian cystatin C assay showed good agreement with the corresponding assay from Dade Behring (Deerfield, Ill., USA) and good inter-laboratory concordance. The assay has very low total imprecision, good linearity and strong correlation with iohexol clearance (R (2) = 0.956). The equation for the correlation curve is: y = 79.901x(-1.4389). CONCLUSIONS: There was low inter-laboratory variation between the three laboratories involved in the cystatin C evaluation, and thus all three laboratories can use the same equation for calculating the estimated GFR.
  •  
Skapa referenser, mejla, bekava och länka
  • Result 1-1 of 1
Type of publication
journal article (1)
Type of content
peer-reviewed (1)
Author/Editor
Larsson, Anders (1)
Hansson, L O (1)
Flodin, Mats (1)
Jonsson, A-S. (1)
Danielsson, L-A. (1)
University
Uppsala University (1)
Language
English (1)
Year

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view