On-Line Biological Sample Cleanup for Electrospray Mass Spectrometry Using Sol-Gel Columns
- 2006
-
Ingår i: Journal of chromatography. B. - : Elsevier BV. - 1570-0232 .- 1873-376X. ; 842:1, s. 70-74
-
Tidskriftsartikel (refereegranskat)abstract
- Using a slight overpressure, a urine sample is loaded onto a monolithic photopolymerized sol-gel column that has been derivatized with hydrophobic carbon chains and then the complex urine matrix is washed with aqueous solution. A buffer containing organic solvent is used to elute the adsorbed peptides by an applied voltage and the sample is then introduced into a mass spectrometer by sheath flow electrospray. The importance of desalting this type of sample is demonstrated by an experiment that shows that the signal intensity of a test solution with neurotensin, sprayed directly into the mass spectrometer, decreased from 4.5 x 10(4) Cps to no detectible signal when just 10% urine is added to the sample solution. We suggest that this procedure may find general application for desalting biological samples prior to mass spectrometric analysis.
|
|