1.
  • Almquist, H, et al. (author)
  • Performance of simultaneous emission-transmission systems for attenuation-corrected SPEct: a method for validation applied to two camera systems
  • 2001
  • In: Nuclear Medicine Communications. - 1473-5628. ; 22:7, s. 759-766
  • Journal article (peer-reviewed)abstract
    • Several commercially available systems for attenuation correction in single photon emission computed tomography (SPECT) based on a transmission scan have been introduced that vary in performance. A test procedure for attenuation correction in SPECT is described and applied to two principally different gamma camera systems (the Siemens Multispect 3 triple-headed system [3HS] and the ADAC Genesys Vertex double-headed system [2HS]). The test procedure was based on geometrically well-defined phantoms. A torso phantom was used to illustrate the attenuation correction methods. The test procedure can be used without detailed knowledge of or access to the algorithms used for attenuation correction. The influence on the transmission measurement of radioactivity in a phantom was higher for the 2HS than for the 3HS. The 3HS produced satisfactory attenuation maps and corrected emission count rates to a constant value independent of phantom density and size. With the 2HS, there was a progressive decrease in the correction of emission count rates with increasing phantom density, and about 30% lower corrected count rates in the large compared with the small phantom. A decrease in measured attenuation coefficients in the vicinity of an emission source was demonstrated in large but not small phantoms. A likely explanation is erroneous correction of downscatter into the transmission energy window. This study demonstrates the need for independent evaluation of systems for attenuation correction in SPECT.
  •  
Skapa referenser, mejla, bekava och länka
  • Result 1-1 of 1
Type of publication
journal article (1)
Type of content
peer-reviewed (1)
Author/Editor
Wollmer, Per (1)
Jonson, Björn (1)
Almquist, H (1)
Norrgren, K (1)
Palmer, J (1)
University
Lund University (1)
Language
English (1)
Research subject (UKÄ/SCB)
Medical and Health Sciences (1)
Year

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Close

Copy and save the link in order to return to this view