- Ask, Per, 1950-, et al.
(författare)
-
Clinical evaluation of different fluid-filled systems for oesophageal manometry
- 1979
-
Ingår i: Scandinavian Journal of Gastroenterology. - 0036-5521 .- 1502-7708. ; 14:1, s. 1-5
-
Tidskriftsartikel (refereegranskat)abstract
- In a clinical study of oesophageal manometry with fluid-filled catheters, both a non-perfused system and a perfused system with a syringe pump have been compared to a system with a low-compliance perfusion pump, which served as a reference. Significantly lower values of motility amplitudes, motility derivatives, and partly of LES pressures, and a time delay of up to 0.5 sec of the amplitude maximum were obtained with the non-perfused system and the system with a syringe pump in comparison to the low-compliance system. Since the oesophageal function can be erroneously evaluated by use of a non-perfused system or a perfused system with a syringe pump, such systems cannot be recommended for clinical use.
|
|