SwePub
Sök i LIBRIS databas

  Utökad sökning

L773:2029 3380
 

Sökning: L773:2029 3380 > On-line Heart Rate ...

On-line Heart Rate Monitoring Using the Extra-corporeal Pressure Sensors of a Dialysis Machine

Holmer, Mattias (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Grigonyte, Egle (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Olde, Bo (författare)
visa fler...
Sternby, Jan (författare)
Segelmark, Mårten (författare)
Lund University,Lunds universitet,Njurmedicin,Sektion II,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Nephrology,Section II,Department of Clinical Sciences, Lund,Faculty of Medicine
Solem, Kristian (författare)
Lund University,Lunds universitet,Institutioner vid LTH,Lunds Tekniska Högskola,Departments at LTH,Faculty of Engineering, LTH
Sörnmo, Leif (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
visa färre...
 (creator_code:org_t)
2012
2012
Engelska.
Ingår i: Biomedical engineering – 2012 : proceedings of international conference. - 2029-3380.
  • Konferensbidrag (refereegranskat)
Abstract Ämnesord
Stäng  
  • Heart rate can be extracted from the extracorporeal venous pressure signal of a dialysis machine. The results are of comparable accuracy and reliability to the ones obtained by the PPG reference signal from a pulse oximeter. Difficulties that occur during heart rate estimation were determined, and some can be overcome by a slight adjustment of blood flow rate. The described techniques, after being implemented into the dialysis machines, would help to improve current hemodialysis safety throughout the treatment by on-line monitoring of the cardiac activity.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)

Publikations- och innehållstyp

kon (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy