Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

Andersson, Karin M., 1989- (författare)
Department of Medical Physics, Örebro University Hospital, Örebro, Sweden
Ahnesjö, Anders (författare)
Uppsala universitet,Enheten för medicinsk strålfysik
Vallhagen Dahlgren, Christina (författare)
Skandionkliniken, Uppsala, Sweden.
 (creator_code:org_t)
2014-09-08
2014
Engelska.
Ingår i: Journal of Applied Clinical Medical Physics. - : John Wiley & Sons. - 1526-9914. ; 15:5, s. 112-119
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Metal objects in the body such as hip prostheses cause artifacts in CT images. When CT images degraded by artifacts are used for treatment planning of radiotherapy, the artifacts can yield inaccurate dose calculations and, for particle beams, erroneous penetration depths. A metal artifact reduction software (O-MAR) installed on a Philips Brilliance Big Bore CT has been tested for applications in treatment planning of proton radiotherapy. Hip prostheses mounted in a water phantom were used as test objects. Images without metal objects were acquired and used as reference data for the analysis of artifact-affected regions outside of the metal objects in both the O-MAR corrected and the uncorrected images. Water equivalent thicknesses (WET) based on proton stopping power data were calculated to quantify differences in the calculated proton beam penetration for the different image sets. The WET to a selected point of interest between the hip prostheses was calculated for several beam directions of clinical relevance. The results show that the calculated differences in WET relative to the reference case were decreased when the O-MAR algorithm was applied. WET differences up to 2.0 cm were seen in the uncorrected case while, for the O-MAR corrected case, the maximum difference was decreased to 0.4 cm. The O-MAR algorithm can significantly improve the accuracy in proton range calculations. However, there are some residual effects, and the use of proton beam directions along artifact streaks should only be used with caution and appropriate margins.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

metal artifact reduction
radiation therapy
water equivalent thickness
proton range

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy