Inadequate white cell reduction by bedside filtration of red cell concentrates

Ledent, Elisabeth (författare)
Linköpings universitet,Östergötlands Läns Landsting,Transfusionsmedicin och klinisk immunologi,Hälsouniversitetet
Berlin, Gösta (författare)
Linköpings universitet,Östergötlands Läns Landsting,Transfusionsmedicin och klinisk immunologi,Hälsouniversitetet
 (creator_code:org_t)
Wiley, 1994
1994
Engelska.
Ingår i: Transfusion. - : Wiley. - 0041-1132 .- 1537-2995. ; 34:9, s. 765-768
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: White cell filtration of red cell concentrates is often performed at the bedside, in the ward, with the filter inserted in the blood administration line. The aim of this study was to evaluate the efficiency of this filtration method and compare it to filtration in the blood bank.Study Design and Methods: One-day-old, buffy coat-reduced, hard-packed red cell concentrates in saline-adenine-glucose-mannitol solution were filtered through different filters designed for bedside or laboratory use. With filters designed for bedside use, filtration of red cells was performed under laboratory conditions at fast flow (10 min) or under bedside conditions at slow flow (2 hours). The remaining white cells were counted microscopically. Filters designed for laboratory use were evaluated at fast flow, and the number of contaminating white cells was counted by flow cytometry.Results: With bedside fllters, a significantly higher contamination of white cells was found In the units filtered at slow flow than at fast flow, regardless of the filter used. The number of units with >5 x 106 white cells was 52 (78%) of 67 filtered at slow flow compared to 11 (23%) of 47 at fast flow, all filters taken together. This difference in white cell contamination was mainly due to an increase of polymorphonuclear cells in the red cell concentrates filtered at slow flow. With filters designed for laboratory use, 0 to 2 percent of units (n = 1448) were contaminated with >5 x 106 white cells.Conclusion: Bedside filtration for white cell reduction at slow flow is inefficient for 1-day-old, buffy coat-reduced red cell concentrates.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Ledent, Elisabet ...
Berlin, Gösta
Artiklar i publikationen
Transfusion
Av lärosätet
Linköpings universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy