SwePub
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Hooker Jonathan)
 

Sökning: WFRF:(Hooker Jonathan) > (2017) > Quantifying Abdomin...

Quantifying Abdominal Adipose Tissue and Thigh Muscle Volume and Hepatic Proton Density Fat Fraction : Repeatability and Accuracy of an MR Imaging–based, Semiautomated Analysis Method

Middleton, Michael (författare)
Department of Radiology, University of California, San Diego, CA, USA
Haufe, William (författare)
Department of Radiology, University of California, San Diego, CA, USA
Hooker, Jonathan (författare)
Department of Radiology, University of California, San Diego, CA, USA
visa fler...
Borga, Magnus (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Medicinsk informatik,Tekniska högskolan
Dahlqvist Leinhard, Olof, 1978- (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Region Östergötland, Radiofysikavdelningen US
Romu, Thobias (författare)
Linköpings universitet,Medicinsk informatik,Tekniska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
Tunón, Patrik (författare)
Advanced MR Analytics AB, Linköping
Hamilton, Gavin (författare)
Department of Radiology, University of California, San Diego, CA
Wolfson, Tanya (författare)
Computational and Applied Statistics Laboratory (CASL), San Diego Supercomputing Center (SDSC), University of California, San Diego, CA
Gamst, Anthony (författare)
Computational and Applied Statistics Laboratory (CASL), San Diego Supercomputing Center (SDSC), University of California, San Diego, CA
Loomba, Rohit (författare)
Department of Medicine (Division of Gastroenterology and Hepatology), University of California, San Diego, CA
Sirlin, Claude (författare)
Department of Radiology, University of California, San Diego, CA
visa färre...
 (creator_code:org_t)
Radiological Society of North America, Inc. 2017
2017
Engelska.
Ingår i: Radiology. - : Radiological Society of North America, Inc.. - 0033-8419 .- 1527-1315. ; 283:2, s. 438-449
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PurposeThe purpose of this study was to determine the repeatability and accuracy of an   commercially available (Advanced MR Analytics [AMRA®]; Linköping, Sweden) magnetic resonance imaging (MRI)-based, semi-automated method to quantify abdominal adipose tissue and thigh muscle volume as well as hepatic proton density fat fraction (PDFF)Materials and MethodsThis prospective study was approved by an institutional review board (IRB) and was Health Insurance Portability and Accountability Act (HIPAA) compliant. All subjects provided written informed consent. Inclusion criteria were age ≥ 18 years, and willingness to participate. Exclusion criteria were contraindication to MRI. Three-dimensional, T1-weighted, dual-echo body-coil images were acquired from base of skull to knees at 3T, twice before and once after taking subjects off the scanner table (total of three acquisitions). Source images were reconstructed offline to generate water, and calibrated fat images where pure adipose tissue has unit value and absence of adipose tissue has zero value. Abdominal adipose tissues and thigh muscles were segmented, and their volumes estimated using AMRA  a semi-automated analysis method and, as a reference standard, manually. Hepatic PDFF was estimated using a confounder-corrected chemical-shift encoded MRI method with hybrid complex-magnitude reconstruction., and, as a reference standard, with magnetic resonance spectroscopy (MRS). Tissue volume and hepatic PDFF intra- and inter-examination repeatability was assessed by intraclass correlation (ICC) and coefficient of variation (CV) analysis. Tissue volume and hepatic PDFF accuracies were assessed by linear regression using their respective reference standards.ResultsTwenty adult subjects were enrolled (18 female, age range 25 - 76 yrs, body mass index range 19.3 to 43.9 kg/m2). Adipose and thigh muscle tissue volumes estimated using the semi-automated analysis method had intra-and inter-examination ICCs between 0.996 and 0.998, and CVs between 1.5 and 3.6%. For hepatic MRI PDFF, intra- and inter-examination ICCs were ≥ 0.994 and CVs, ≤ 7.3%. Agreement between semi-automated and manual volume estimates, and between MRI and MRS hepatic PDFF estimates, was high, with regression slopes and intercepts not significantly different from the identity line (all p’s > 0.05), and R2’s between 0.744 and 0.994.ConclusionsThis MRI-based, semi-automated method provides high repeatability, and high accuracy for estimating abdominal adipose tissue and thigh muscle volumes, and hepatic PDFF.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Nyckelord

Liver
proton density fat fraction
PDFF
repeatability
accuracy
semi-automated
SCAT
VAT.

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

  • Radiology (Sök värdpublikationen i LIBRIS)

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy