Reduction of motion, truncation and flow artifacts using BLADE sequences in cervical spine MR imaging

Lavdas, Eleftherios (författare)
Mavroidis, Panayiotis (författare)
Stockholms universitet,Medicinsk strålningsfysik (tills m KI),University of Texas Health Sciences Center at San Antonio, USA
Kostopoulos, Spiros (författare)
visa fler...
Ninos, Constantin (författare)
Strikou, Aspasia-Dimitra (författare)
Glotsos, Dimitrios (författare)
Vlachopoulou, Anna (författare)
Oikonomou, Georgia (författare)
Economopoulos, Nikolaos (författare)
Roka, Violeta (författare)
Sakkas, Georgios K. (författare)
Tsagkalis, Antonios (författare)
Stathakis, Sotirios (författare)
Papanikolaou, Nikos (författare)
Batsikas, Georgios (författare)
visa färre...
 (creator_code:org_t)
Elsevier BV, 2015
2015
Engelska.
Ingår i: Magnetic Resonance Imaging. - : Elsevier BV. - 0730-725X .- 1873-5894. ; 33:2, s. 194-200
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose: To assess the efficacy of the BLADE technique (MR imaging with 'rotating blade-like k-space covering') to significantly reduce motion, truncation, flow and other artifacts in cervical spine compared to the conventional technique.Materials and methods: In eighty consecutive subjects, who had been routinely scanned for cervical spine examination, the following pairs of sequences were compared: a) T2 TSE SAG vs. T2 TSE SAG BLADE and b) T2 TIRM SAG vs. T2 TIRM SAG BLADE. A quantitative analysis was performed using the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) measures. A qualitative analysis was also performed by two radiologists, who graded seven image characteristics on a 5-point scale (0: non-visualization; 1: poor; 2: average; 3: good; 4: excellent). The observers also evaluated the presence of image artifacts (motion, truncation, flow, indentation).Results: In quantitative analysis, the CNR values of the CSF/SC between TIRM SAG and TIRM SAG BLADE were found to present statistically significant differences (p < 0.001). Regarding motion and truncation artifacts, the T2 TSE BLADE SAG was superior compared to the T2 TSE SAG, and the T2 TIRM BLADE SAG was superior compared to the T2 TIRM SAG. Regarding flow artifacts, T2 TIRM BLADE SAG eliminated more artifacts than T2 TIRM SAG.Conclusions: In cervical spine MRI, BLADE sequences appear to significantly reduce motion, truncation and flow artifacts and improve image quality. BLADE sequences are proposed to be used for uncooperative subjects. Nevertheless, more research needs to be done by testing additional specific pathologies.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Cervical spine MR imaging
BLADE sequences
Image quality
Artifact reduction

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy