Allergic contact dermatitis caused by isobornyl acrylate in the Enlite glucose sensor and the Paradigm MiniMed Quick-set insulin infusion set
-
- Herman, Anne (författare)
- Saint-Luc University Hospital
-
- Baeck, Marie (författare)
- Saint-Luc University Hospital
-
- de Montjoye, Laurence (författare)
- Saint-Luc University Hospital
-
visa fler...
-
- Bruze, Magnus (författare)
- Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
-
- Giertz, Emil (författare)
- Blekinge Hospital
-
- Goossens, An (författare)
- University Hospitals Leuven
-
- Mowitz, Martin (författare)
- Lund University,Lunds universitet,Yrkes- och miljödermatologi,Forskargrupper vid Lunds universitet,Occupational and Environmental Dermatology,Lund University Research Groups,Skåne University Hospital
-
visa färre...
-
(creator_code:org_t)
- 2019-08-27
- 2019
- Engelska.
-
Ingår i: Contact Dermatitis. - : Wiley. - 0105-1873 .- 1600-0536. ; 81:6, s. 432-437
- Relaterad länk:
-
http://dx.doi.org/10...
-
visa fler...
-
https://lup.lub.lu.s...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Background: The FreeStyle Libre glucose sensor has caused many cases of allergic contact dermatitis, and isobornyl acrylate (IBOA) in this sensor has been identified as one of the culprit allergens. Objectives: To report on the presence of IBOA in devices produced by Medtronic, namely, Enlite sensor and insulin infusion set Paradigm MiniMed Quick-set. Patients and Methods: Five patients reacting to the glucose sensor Enlite and/or the insulin infusion set Paradigm MiniMed Quick-set observed in three clinics (two Belgian and one Swedish) were patch tested with the baseline and other series, as well as with IBOA; four of them also with pieces of adhesive patches from the devices, and two with a thin layer chromatogram of Enlite glucose sensor extracts. Gas chromatography-mass spectrometry (GC-MS) analyses were performed. Results: Four patients reacted to IBOA and one to colophonium, a known allergen in Enlite, and three to the adhesive part of the sensor or the insulin infusion set. IBOA was identified in the sensor by GC-MS, and its presence was indicated in the infusion set. Conclusions: IBOA is a contact allergen in Enlite glucose sensor, and likely also in the infusion set. Therefore, these devices are not suitable alternatives for patients sensitized to the FreeStyle Libre sensor.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Dermatologi och venereologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Dermatology and Venereal Diseases (hsv//eng)
Nyckelord
- acrylates
- allergic contact dermatitis
- delayed hypersensitivity
- glucose sensor
- insulin pump
- isobornyl acrylate
- medical device