The choice of anti-scatter grids in diagnostic radiology : the optimization of image quality and absorbed dose

Sandborg, Michael (författare)
Östergötlands Läns Landsting,Linköpings universitet,Radiofysik,Hälsouniversitetet,Radiofysikavdelningen
Dance, David R. (författare)
Department of Medical Physics, The Royal Marsden Hospital, Fulham Road, SW3 6JJ London, United Kingdom
Alm Carlsson, Gudrun (författare)
Östergötlands Läns Landsting,Linköpings universitet,Radiofysik,Hälsouniversitetet,Radiofysikavdelningen
visa fler...
Persliden, Jan (författare)
Östergötlands Läns Landsting,Linköpings universitet,Radiofysik,Hälsouniversitetet,Radiofysikavdelningen
visa färre...
 (creator_code:org_t)
Linköping : Linköping University Electronic Press, 1993
Engelska 40 s.
Serie: Report / Institutionen för radiofysik, Universitetet i Linköping, 1990-1997, 1102-1799 ; 72
Serie: SSI-rapport / Statens strålskyddsinstitut - SSI ; Swedish Radiation Protection Institute, 0282-4434 ; SSI P 586:90
  • Rapport (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • A Monte Carlo model is developed to study and optimise the design of antiscatter grids in diagnostic radiology. The imaging chain including X-ray energy spectra, phantom (representing the patient), grid and image receptor is simulated. Image quality is quantified in terms of contrast (conventionai screen-film imaging) and signal-to-noise ratio, SNR (digital imaging) and the radiation detriment to the patient (risk) by the mean absorbed dose in the phantom. The advantages of using fibre instead of aluminium for grid interspaces and covers are quantified. Compared to aluminium grids, the absorbed dose is reduced by 10-50%, contrast is improved by 0-10% and SNR by 10-40% (digital radiography). The advantages are larger at low tube potentials and for grids with high ratio and low strip density. Commercial grids, with different interspace materials, strip density, strip width and grid ratio, are compared in paediatric, lumbar spine and chest examinations. The differences in dose increase and contrast improvement factors obtained with these grids are mainly due to the use of different materials in the grid interspaces, but the strip design is also important. In a global optimisation of grid design and tube potential at fixed contrast, it is found that grids of different strip density and ratio all can have good performances provided that they are used with appropriate strip width and tube potential. In the paediatric examination, low ratio grids need thinner strips than used today to be optimal. A small air gap could alternatively be used. In examinations with more scatter (adult AP), present commercial grids are optimal (r=12-16, d=30-50µm). In the lateral view (even more scatter), grids with ratios larger than 16 are optimal provided the grid can be accurately aligned in the beam. The optimization is performed with grids with fibre interspaces and covers since low atomic number materials should preferably be used for materials between the patient and the receptor. Optimal grids with aluminium for these components have lower grid ratio and higher strip densities than optimal fibre grids.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Radiological physics
Radiofysik

Publikations- och innehållstyp

vet (ämneskategori)
rap (ämneskategori)

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy