- Herrlin, K, et al.
(författare)
-
Generation of X-rays For Medical Imaging By High-power Lasers - Preliminary-results
- 1993
-
Ingår i: Radiology. - 1527-1315. ; 189:1, s. 65-68
-
Tidskriftsartikel (refereegranskat)abstract
- PURPOSE: To evaluate the use of x-ray imaging performed with a high-power laser system in biologic and medical studies. MATERIALS AND METHODS: A compact terawatt laser system based on chirped pulse amplification in titanium-doped sapphire was used. X rays were generated by irradiating a high-atomic-number target (tantalum). RESULTS: When photons with an energy below 10 keV were removed with use of 3 mm of aluminum, the half-value layer in aluminum for the remaining x rays was approximately 10 mm. The x-ray source allowed performance of biologic magnification radiography. Star-pattern tests indicated an equivalent focal spot size of less than 60 mum. Exposures of a single pulse could be obtained. The duration of a single x-ray pulse was estimated to be of the order of picoseconds. CONCLUSION. With use of subpicosecond laser pulses, x-ray generation can occur with a smaller equivalent focal spot size than with conventional x-ray sources.
|
|