- Goryashko, Vitaliy, et al.
(författare)
-
HIGH-PRECISION MEASUREMENTS OF THE QUALITY FACTOR OFSUPER CONDUCTING CAVITIES AT THE FREIA LABORATORY
- 2015
-
Konferensbidrag (refereegranskat)abstract
- The dependence of cavity quality factor Q0 on accelerating gradient gives insight into the intrinsic limit of RFsurface impedance that determines the cavity performance. In this paper we propose a high-precision method of measuringQ0 of SRF cavities. A common way to study the performance of an SRF cavity is to build an oscillator around it that is referred to as a self-excited loop. In the standard approach, by tuning the loop phase for a maximum field level in thecavity and measuring forward and reflected waves, one finds the cavity coupling. Then, performing a time-decay measurement and finding the total quality factor, one gets Q0. However, this approach suffers from a deficiency originating from a single data-point measurement of the reflection coefficient. In our method by varying the loop phase shift, one obtains amplitudes of the reflection coefficient of the cavity as a function of its phases. The complex reflection coefficient describes a perfect circle in polar coordinates. Fitting the overdetermined set of data to that circle allows more accurate calculation of Q0 via the least-squares procedure. The method has been tested at the FREIA Laboratory on two cavities from IPN Orsay: a single spoke and a prototype ESS double spoke.
|
|