SwePub
Tyck till om SwePub Sök här!
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "WFRF:(Grindborg Jan Erik) "

Sökning: WFRF:(Grindborg Jan Erik)

  • Resultat 1-7 av 7
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  •  
2.
  • Adolfsson, Emelie, 1985-, et al. (författare)
  • Response of Lithium Formate EPR Dosimeters at Photon Energies Relelvant to Brachytherapy
  • 2009
  • Ingår i: IFMBE Proceedings. - Heidelberg : Springer Berlin Heidelberg. - 9783642034725 - 9783642034749 ; , s. 236-239
  • Konferensbidrag (övrigt vetenskapligt/konstnärligt)abstract
    • After development of sensitive dosimeter materials Electron Paramagnetic Resonance EPR dosimetry has been successfully used also in radiation therapy. The intensity of the EPR-signal is a measure of the amount of free radicals created by ionizing radiation which is proportional to the absorbed dose in the dosimeter. Lithium formate monohydrate is a dosimeter material with 2-6 times higher sensitivity than alanine, a linear dose response over a wide dose range and mass-energy absorption properties similar to water. These properties make lithium formate promising for verification of absorbed doses around high dose rate brachytherapy sources where the dose gradient is steep and the photon energy distribution changing with distance from the source. Calibration of the dosimeters is performed in 60Co or MV photon beams where high dosimetric accuracy is feasible. The use in brachytherapy field relies on the assumption that the production of free radicals per mean absorbed dose in the dosimeter is similar at the lower photon energies present there. The aim of this work was to test that assumption. The response of the dosimeters as a function of photon energy was determined by irradiations with four x-ray qualities in the range 100-250 kV and 137Cs, relative to the response when irradiated with 60Co, all photon beams with well-known air kerma rates at the Swedish Secondary Standards Dosimetry Laboratory. Monte Carlo simulations were used to convert air kerma free in air to mean absorbed dose to the dosimeter. The measured response relative 60Co as a function of photon energy was below unity for all qualities. The maximum deviation from unity was 2.5% (100 kV, 135 kV) with a relative standard deviation of 1.5% (k = 1).
  •  
3.
  • Adolfsson, Emelie, et al. (författare)
  • Response of lithium formate EPR dosimeters at photon energies relevant to the dosimetry of brachytherapy
  • 2010
  • Ingår i: Medical physics (Lancaster). - : American Association of Physicists in Medicine. - 0094-2405. ; 37:9, s. 4946-4959
  • Tidskriftsartikel (refereegranskat)abstract
    • PURPOSE:To investigate experimentally the energy dependence of the detector response of lithium formate EPR dosimeters for photon energies below 1 MeV relative to that at 60Co energies. High energy photon beams are used in calibrating dosimeters for use in brachytherapy since the absorbed dose to water can be determined with high accuracy in such beams using calibrated ion chambers and standard dosimetry protocols. In addition to any differences in mass-energy absorption properties between water and detector, variations in radiation yield (detector response) with radiation quality, caused by differences in the density of ionization in the energy imparted (LET), may exist. Knowledge of an eventual deviation in detector response with photon energy is important for attaining high accuracy in measured brachytherapy dose distributions.METHODS:Lithium formate EPR dosimeters were irradiated to known levels of air kerma in 25-250 kV x-ray beams and in 137Cs and 60Co beams at the Swedish Secondary Standards Dosimetry Laboratory. Conversions from air kerma free in air into values of mean absorbed dose to the detectors were made using EGSnrc MC simulations and x-ray energy spectra measured or calculated for the actual beams. The signals from the detectors were measured using EPR spectrometry. Detector response (the EPR signal per mean absorbed dose to the detector) relative to that for 60Co was determined for each beam quality.RESULTS:Significant decreases in the relative response ranging from 5% to 6% were seen for x-ray beams at tube voltages < or = 180 kV. No significant reduction in the relative response was seen for 137Cs and 250 kV x rays.CONCLUSIONS:When calibrated in 60Co or MV photon beams, corrections for the photon energy dependence of detector response are needed to achieve the highest accuracy when using lithium formate EPR dosimeters for measuring absorbed doses around brachytherapy sources emitting photons in the energy range of 20-150 keV such as 169Yb and electronic sources.
  •  
4.
  • Carlsson Tedgren, Åsa, 1968-, et al. (författare)
  • Audit on source strength determination for HDR and PDR 192Ir brachytherapy in Sweden
  • 2008
  • Ingår i: Radiotherapy and Oncology. - : Elsevier BV. - 0167-8140 .- 1879-0887. ; 86:1, s. 126-130
  • Tidskriftsartikel (refereegranskat)abstract
    • Background and purpose: To investigate the status of source strength determination in terms of reference air kerma rate (RAKR) for HDR and PDR 192Ir brachytherapy in Sweden. Materials and methods: RAKR was determined in each of the 14 Swedish afterloaders, using calibrated equipment from the Swedish Secondary Standard Dosimetry Laboratory. Results: Values of RAKR from the external audit, the hospitals and vendors agreed within the uncertainty limits guaranteed by the vendors. Conclusions: The accuracy in RAKR determination has increased over the last years as a result of increased availability of interpolation standards for HDR 192Ir and the increased use of robust well-type ion chambers designed for brachytherapy. It is recommended to establish a ratio between the RAKR value from own measurements at the hospital and that of the vendor since such a ratio embeds constant systematic differences due to e.g. varying traceability and therefore has the potential of being less uncertain than the RAKR alone. Traceability to primary standards for HDR 192Ir sources will in the future significantly decrease the uncertainty in RAKR of 192Ir brachytherapy. © 2007 Elsevier Ireland Ltd. All rights reserved.
  •  
5.
  • Carlsson Tedgren, Åsa, et al. (författare)
  • Characterization of a Co-60 unit at a secondary standard dosimetry laboratory: Monte Carlo simulations compared to measurements and results from the literature
  • 2010
  • Ingår i: Medical physics (Lancaster). - : American Association of Physicists in Medicine. - 0094-2405. ; 37:6, s. 2777-2786
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: To compare a Monte Carlo (MC) characterization of a Co-60 unit at the Swedish Secondary Standard Dosimetry Laboratory (SSDL) with the results of both measurements and literature with the aims of (1) resolving a change in the ratio of air-kerma free in air K-air and absorbed dose to water D-w in a water phantom noted experimentally after a source exchange in the laboratory and (2) reviewing results from the literature on similar MC simulations. Although their use in radiotherapy is decreasing, the characteristics of Co-60 beams are of interest since Co-60 beams are utilized in calibrating ionization chambers for the absolute dosimetry of radiotherapy beams and as reference radiation quality in evaluating the energy dependence of radiation detectors and in studies on radiobiological effectiveness. Methods: The BEAMnrc MC code was used with a detailed geometrical model of the treatment head and two models of the Co-60 source representing the sources used before and after source exchange, respectively. The active diameters of the Co-60 sources were 1.5 cm in pellet form and 2.0 cm in sintered form. Measurements were performed on the actual unit at the Swedish SSDL. Results: Agreement was obtained between the MC and the measured results within the estimated uncertainties for beam profiles, water depth-dose curve, relative air-kerma output factors, and for the ratios of K-air/D-w before and after source exchange. The on-axis energy distribution of the photon fluence free in air for the unit loaded with its present (1.5 cm in diameter) source agreed closely with the results from the literature in which a source of the same make and active diameter, inside a different treatment head, was simulated. The spectrum for the larger (2.0 cm in diameter) source was in close agreement with another published spectrum, also modeling a Co-60 source with an active diameter of 2.0 cm inside a different treatment head. Conclusions: The reduction in the value of K-air/D-w following source exchange was explained by the spectral differences between the two sources that were larger in the free in-air geometry used for K-air calibrations than at 5 g/cm(2) depth in the water phantom used for D-w calibrations. Literature review revealed differences between published in-air Co-60 spectra derived for sources of different active diameters, and investigators in need of an accurately determined Co-60 in-air spectrum should be aware of differences due to source active diameter.
  •  
6.
  • Carlsson Tedgren, Åsa, et al. (författare)
  • Comparison of high-dose-rate Ir-192 source strength measurements using equipment with traceability to different standards
  • 2014
  • Ingår i: Brachytherapy. - : Elsevier. - 1538-4721 .- 1873-1449. ; 13:4, s. 420-423
  • Tidskriftsartikel (refereegranskat)abstract
    • PURPOSE:According to the American Association of Physicists in Medicine Task Group No. 43 (TG-43) formalism used for dose calculation in brachytherapy treatment planning systems, the absolute level of absorbed dose is determined through coupling with the measurable quantity air-kerma strength or the numerically equal reference air-kerma rate (RAKR). Traceability to established standards is important for accurate dosimetry in laying the ground for reliable comparisons of results and safety in adoption of new treatment protocols. The purpose of this work was to compare the source strength for a high-dose rate (HDR) (192)Ir source as measured using equipment traceable to different standard laboratories in Europe and the United States.METHODS AND MATERIALS:Source strength was determined for one HDR (192)Ir source using four independent systems, all with traceability to different primary or interim standards in the United States and Europe.RESULTS:The measured HDR (192)Ir source strengths varied by 0.8% and differed on average from the vendor value by 0.3%. Measurements with the well chambers were 0.5% ± 0.1% higher than the vendor-provided source strength. Measurements with the Farmer chamber were 0.7% lower than the average well chamber results and 0.2% lower than the vendor-provided source strength. All of these results were less than the reported source calibration uncertainties (k=2) of each measurement system.CONCLUSIONS:In view of the uncertainties in ion chamber calibration factors, the maximum difference in source strength found in this study is small and confirms the consistency between calibration standards in use for HDR (192)Ir brachytherapy.
  •  
7.
  • Carlsson Tedgren, Åsa, et al. (författare)
  • Response of LiF:Mg,Ti thermoluminescent dosimeters at photon energies relevant to the dosimetry of brachytherapy (andlt; 1 MeV)
  • 2011
  • Ingår i: Medical physics (Lancaster). - : American Association of Physicists in Medicine. - 0094-2405. ; 38:10, s. 5539-5550
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: High energy photon beams are used in calibrating dosimeters for use in brachytherapy since absorbed dose to water can be determined accurately and with traceability to primary standards in such beams, using calibrated ion chambers and standard dosimetry protocols. For use in brachytherapy, beam quality correction factors are needed, which include corrections for differences in mass energy absorption properties between water and detector as well as variations in detector response (intrinsic efficiency) with radiation quality, caused by variations in the density of ionization (linear energy transfer (LET) -distributions) along the secondary electron tracks. The aim of this work was to investigate experimentally the detector response of LiF:Mg, Ti thermoluminescent dosimeters (TLD) for photon energies below 1 MeV relative to (60)Co and to address discrepancies between the results found in recent publications of detector response. less thanbrgreater than less thanbrgreater thanMethods: LiF:Mg,Ti dosimeters of formulation MTS-N Poland were irradiated to known values of air kerma free-in-air in x-ray beams at tube voltages 25-250 kV, in (137)Cs- and (60)Co-beams at the Swedish Secondary Standards Dosimetry Laboratory. Conversions from air kerma free-in-air into values of mean absorbed dose in the dosimeters in the actual irradiation geometries were made using EGSnrc Monte Carlo simulations. X-ray energy spectra were measured or calculated for the actual beams. Detector response relative to that for (60)Co was determined at each beam quality. less thanbrgreater than less thanbrgreater thanResults: An increase in relative response was seen for all beam qualities ranging from 8% at tube voltage 25 kV (effective energy 13 keV) to 3%-4% at 250 kV (122 keV effective energy) and (137)Cs with a minimum at 80 keV effective energy (tube voltage 180 kV). The variation with effective energy was similar to that reported by Davis [Radiat. Prot. Dosim. 106, 33-43 (2003)] with our values being systematically lower by 2%-4%. Compared to the results by Nunn [Med. Phys. 35, 1861-1869 (2008)], the relative detector response as a function of effective energy differed in both shape and magnitude. This could be explained by the higher maximum read-out temperature (350 degrees C) used by Nunn [Med. Phys. 35, 1861-1869 (2008)], allowing light emitted from high-temperature peaks with a strong LET dependence to be registered. Use of TLD-100 by Davis [Radiat. Prot. Dosim. 106, 33-43 (2003)] with a stronger super-linear dose response compared to MTS-N was identified as causing the lower relative detector response in this work. less thanbrgreater than less thanbrgreater thanConclusions: Both careful dosimetry and strict protocols for handling the TLDs are required to reach solid experimental data on relative detector response. This work confirms older findings that an over-response relative to (60)Co exists for photon energies below 200-300 keV. Comparison with the results from the literature indicates that using similar protocols for annealing and read-out, dosimeters of different makes (TLD-100, MTS-N) differ in relative detector response. Though universality of the results has not been proven and further investigation is needed, it is anticipated that with the use of strict protocols for annealing and read-out, it will be possible to determine correction factors that can be used to reduce uncertainties in dose measurements around brachytherapy sources at photon energies where primary standards for absorbed dose to water are not available.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-7 av 7

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy