A 4-afc study comparing ultrasound machines using a greyscale phantom
-
- Lorentsson, Robert, 1969 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
-
Hosseini, N. (författare)
-
Johansson, J.O. (författare)
-
visa fler...
-
Rosenberg, W. (författare)
-
Stenborg, B. (författare)
-
- Månsson, Lars Gunnar, 1951 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
-
- Båth, Magnus, 1974 (författare)
- Gothenburg University,Göteborgs universitet,Institutionen för kliniska vetenskaper, Avdelningen för radiofysik,Institute of Clinical Sciences, Department of Radiation Physics
-
visa färre...
-
(creator_code:org_t)
- Cham : Springer International Publishing, 2015
- 2015
- Engelska.
-
Ingår i: IFMBE Proceedings. 16th Nordic-Baltic Conference on Biomedical Engineering. October 14-16, 2014, Gothenburg, Sweden. - Cham : Springer International Publishing. - 1680-0737. ; 48, s. 82-85
- Relaterad länk:
-
https://gup.ub.gu.se...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- © Springer International Publishing Switzerland 2015. Two ultrasound machines, one high-end and one ordinary, were compared in terms of their ability to reproduce low-contrast objects using the same probe. Images containing 4 mm objects of four different contrasts were collected from a greyscale phantom at 35-42 mm depth. Six observers participated in a 4-alternative forced choice study based on 120 images. At this certain depth and object size the proportion of correct responses was higher (statistically significant) for the high-end machine at three of four contrast levels, indicating the possibility to discriminate between ultrasound machines using a limited number of images of a greyscale phantom. However, the number of images and number of observers needed are larger than usually used for constancy control.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- 4-AFC
- Detection
- Phantom observation
- Ultrasound