Cole Function and Conductance-Based Parasitic Capacitance Compensation for Cerebral Electrical Bioimpedance Measurements
-
- Atefi, Seyed Reza, 1985- (författare)
- KTH,Medicinska sensorer, signaler och system (MSSS)
-
- Buendia, Ruben, 1983- (författare)
- Högskolan i Borås,KTH,Medicinska sensorer, signaler och system (MSSS),Institutionen Ingenjörshögskolan
-
- Lindecrantz, Kaj, 1951- (författare)
- Högskolan i Borås,Karolinska Institutet,KTH,Medicinska sensorer, signaler och system (MSSS),Institutionen Ingenjörshögskolan
-
visa fler...
-
- Seoane, Fernando, 1976- (författare)
- Högskolan i Borås,KTH,Medicinska sensorer, signaler och system (MSSS),Institutionen Ingenjörshögskolan
-
visa färre...
-
(creator_code:org_t)
- San Diego : IEEE press, 2012
- 2012
- Engelska.
-
Ingår i: Engineering in Medicine and Biology Society (EMBC), 2012 Annual International Conference of the IEEE. - San Diego : IEEE press. ; 2012, s. 3368-71
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
http://kipublication...
-
https://urn.kb.se/re...
-
visa färre...
Abstract
Ämnesord
Stäng
- One of the most common measurement artifacts present in Electrical Bioimpedance Spectroscopy measurements (EBIS) comes from the capacitive leakage effect resulting from parasitic stray capacitances. This artifact produces a deviation in the measured impedance spectrum that is most noticeable at higher frequencies. The artifact taints the spectroscopy measurement increasing the difficulty of producing reliable EBIS measurements at high frequencies. In this work, an approach for removing such capacitive influence from the spectral measurement is presented making use of a novel method to estimate the value of the parasitic capacitance equivalent that causes the measurement artifact. The proposed method has been tested and validated theoretically and experimentally and it gives a more accurate estimation of the value of the parasitic capacitance than the previous methods. Once a reliable value of parasitic capacitance has been estimated the capacitive influence can be easily compensated in the EBIS measured data. Thus enabling analysis of EBIS data at higher frequencies, i.e. in the range of 300-500 kHz like measurements intended for cerebral monitoring, where the characteristic frequency is remarkably higher than EBIS measurements i.e. within the range 30 to 50 kHz, intended for body composition assessment.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Elektroteknik och elektronik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Electrical Engineering, Electronic Engineering, Information Engineering (hsv//eng)
Nyckelord
- Accurate estimation
- Body composition
- Cerebral monitoring
- Characteristic frequencies
- Electrical bio-impedance
- High frequency
- Higher frequencies
- Leakage effects
- Measured impedance
- Measurement artifacts
- Parasitic capacitance
- Spectral measurement
- Spectroscopy measurements
- Stray capacitances