Sökning: onr:"swepub:oai:DiVA.org:kth-17613" >
Assessment of left ...
Assessment of left ventricular volumes using simplified 3-D echocardiography and computed tomography - a phantom and clinical study
-
- Mårtensson, Mattias (författare)
- KTH,Medicinsk bildteknik
-
- Winter, Reidar (författare)
- Karolinska Institutet
-
- Cederlund, Kerstin (författare)
- Karolinska Institutet
-
visa fler...
-
- Ripsweden, Jonaz (författare)
- Karolinska Institutet
-
- Mir-Akbari, Habib (författare)
- Karolinska Institutet
-
- Nowak, Jacek (författare)
- Karolinska Institutet
-
- Brodin, Lars-Åke (författare)
- Karolinska Institutet,KTH,Medicinsk teknik
-
visa färre...
-
(creator_code:org_t)
- 2008-06-04
- 2008
- Engelska.
-
Ingår i: Cardiovascular Ultrasound. - : Springer Science and Business Media LLC. - 1476-7120. ; 6
- Relaterad länk:
-
https://cardiovascul...
-
visa fler...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
http://kipublication...
-
visa färre...
Abstract
Ämnesord
Stäng
- Objectives: To compare the accuracy of simplified 3-dimensional (3-D) echocardiography vs. multi-slice computed tomography (MSCT) software for the quantification of left ventricular (LV) volumes. Design: Three-D echocardiography (3-planes approach) and MSCT-CardIQ software were calibrated by measuring known volumes of 10 phantoms designed to closely mimic blood-endocardium interface. Subsequently, LV volumes were measured with both the methods in 9 patients referred routinely for coronary angiography and the agreement between the measurements was evaluated. Results: Simplified 3D-echocardiography provided higher degree of agreement between the measured and true phantom volumes (mean difference 0 +/- 1 ml, variation range + 4 to -4 ml) than MSCT software (mean difference 6 +/- 5 ml; variation range + 22 to -10 ml). The agreement between LV measurements in the patients was considerably poorer, with significantly larger volumes produced by MSCT (mean difference - 23 +/- 40 ml, variation between + 93 and -138 ml). Conclusion: Simplified 3-D echocardiography provides more accurate assessment of phantom volumes than MSCT-CardIQ software. The discrepancy between the results of LV measurements with the two methods is even greater and does not warrant their interchangeable diagnostic use.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk apparatteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Equipment Engineering (hsv//eng)
Nyckelord
- cardiac magnetic-resonance
- 3-dimensional echocardiography
- 2-dimensional echocardiography
- ejection fraction
- size
- cineventriculography
- quantification
- transducer
- spect
- Medical engineering
- Medicinsk teknik
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas