Assessment of image quality in abdominal computed tomography : Effect of model-based iterative reconstruction, multi-planar reconstruction and slice thickness on potential dose reduction

Kataria, Bharti, 1955- (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Röntgenkliniken i Linköping
Nilsson Althen, Jonas, 1969- (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Region Östergötland, Medicinsk strålningsfysik
Smedby, Örjan, Professor, 1956- (författare)
KTH,Medicinsk avbildning,KTH Royal Inst Technol, Sweden
visa fler...
Persson, Anders, 1953- (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Röntgenkliniken i Linköping
Sökjer-Petersen, Hannibal (författare)
Linköpings universitet,Medicinska fakulteten,Avdelningen för diagnostik och specialistmedicin
Sandborg, Michael, 1961- (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Medicinsk strålningsfysik
visa färre...
 (creator_code:org_t)
Elsevier Ireland Ltd, 2020
2020
Engelska.
Ingår i: European Journal of Radiology. - : Elsevier Ireland Ltd. - 0720-048X .- 1872-7727. ; 122
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose: To determine the effect of tube load, model-based iterative reconstruction (MBIR) strength and slice thickness in abdominal CT using visual comparison of multi-planar reconstruction images. Method: Five image criteria were assessed independently by four radiologists on two data sets at 42- and 98-mAs tube loads for 25 patients examined on a 192-slice dual-source CT scanner. Effect of tube load, MBIR strength, slice thickness and potential dose reduction was estimated with Visual Grading Regression (VGR). Objective image quality was determined by measuring noise (SD), contrast-to-noise (CNR) ratio and noise-power spectra (NPS). Results: Comparing 42- and 98-mAs tube loads, improved image quality was observed as a strong effect of log tube load regardless of MBIR strength (p < 0.001). Comparing strength 5 to 3, better image quality was obtained for two criteria (p < 0.01), but inferior for liver parenchyma and overall image quality. Image quality was significantly better for slice thicknesses of 2mm and 3mm compared to 1mm, with potential dose reductions between 24%-41%. As expected, with decrease in slice thickness and algorithm strength, the noise power and SD (HU-values) increased, while the CNR decreased. Conclusion: Increasing slice thickness from 1 mm to 2 mm or 3 mm allows for a possible dose reduction. MBIR strength 5 shows improved image quality for three out of five criteria for 1 mm slice thickness. Increasing MBIR strength from 3 to 5 has diverse effects on image quality. Our findings do not support a general recommendation to replace strength 3 by strength 5 in clinical abdominal CT protocols. However, strength 5 may be used in task-based protocols.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Nyckelord

Computed tomography
Abdomen
Iterative reconstruction
Dose
Slice thickness
Multi-planar reconstruction (MPR)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy