Sökning: onr:"swepub:oai:DiVA.org:kth-296401" >
On the Development ...
On the Development of a Novel Contrast Pulse Sequence for Polymer-Shelled Microbubbles
-
- Chen, Hongjian (författare)
- KTH,Medicinsk avbildning
-
- Evangelou, Dimitrios (författare)
- KTH,Medicinteknik och hälsosystem,Dept Biomed Engn & Hlth Syst, S-14152 Stockholm, Sweden.
-
- Loskutova, Ksenia (författare)
- KTH,Medicinsk avbildning
-
visa fler...
-
- Ghorbani, Morteza (författare)
- KTH,Medicinsk avbildning,Sabanci Univ, Nanotechnol Res & Applicat Ctr, TR-34956 Istanbul, Turkey.
-
- Grishenkov, Dmitry, Associate Professor, 1983- (författare)
- KTH,Medicinsk avbildning,Karolinska Inst, Dept Clin Sci Intervent & Technol, S-14152 Stockholm, Sweden.;Karolinska Univ Hosp, Dept Clin Physiol, S-17177 Stockholm, Sweden.
-
visa färre...
-
(creator_code:org_t)
- Institute of Electrical and Electronics Engineers (IEEE), 2021
- 2021
- Engelska.
-
Ingår i: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control. - : Institute of Electrical and Electronics Engineers (IEEE). - 0885-3010 .- 1525-8955. ; 68:5, s. 1569-1579
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
http://kipublication...
-
visa färre...
Abstract
Ämnesord
Stäng
- Contrast agents are routinely used in ultrasound examinations. Nonlinear ultrasound imaging techniques have been developed over decades to enhance the contrast between the tissue and the blood pool after the injection of ultrasound contrast agents (UCAs). In this study, we introduce a new contrast pulse sequence, CPS4. The CPS4 combines pulse inversion (PI), subharmonic (SH), and ultraharmonic (UH) techniques to remove propagation distortion while capturing the unique SH and UH responses from UCAs. The novel CPS4 and conventional PI, SH, and UH techniques were used to detect the presence of a research-grade, thick-shell, polymer microbubble in a tissue-mimicking flow phantom. The contrast-to-tissue ratios (CTRs) obtained from the applications of all techniques were compared. The results show that the highest CTR of approximately 16 dB was obtained using CPS4, which was superior to the individual reference techniques: PI, SH, and UH techniques, in all scenarios considered in this study.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)
Nyckelord
- Ultrasonic imaging
- Imaging
- Transducers
- Finite impulse response filters
- Band-pass filters
- Phantoms
- Optical filters
- Contrast pulse sequence
- medical imaging
- ultrasound contrast agents (UCAs)