Comparing hepatic 2D and 3D magnetic resonance elastography methods in a clinical setting – Initial experiences

Forsgren, Mikael (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Region Östergötland, Radiofysikavdelningen US,Wolfram MathCore AB, Linköping, Sweden
Norén, Bengt (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Region Östergötland, Röntgenkliniken i Linköping
Kihlberg, Johan, 1970- (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Region Östergötland, Röntgenkliniken i Linköping
visa fler...
Dahlqvist Leinhard, Olof (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Region Östergötland, Radiofysikavdelningen US
Kechagias, Stergios (författare)
Linköpings universitet,Avdelningen för kardiovaskulär medicin,Hälsouniversitetet,Region Östergötland, Magtarmmedicinska kliniken
Lundberg, Peter (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Region Östergötland, Radiofysikavdelningen US
visa färre...
 (creator_code:org_t)
Elsevier, 2015
2015
Engelska.
Ingår i: European Journal of Radiology Open. - : Elsevier. - 2352-0477. ; 2, s. 66-70
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PurposeContinuous monitoring of liver fibrosis progression in patients is not feasible with the current diagnostic golden standard (needle biopsy). Recently, magnetic resonance elastography (MRE) has emerged as a promising method for such continuous monitoring. Since there are different MRE methods that could be used in a clinical setting there is a need to investigate whether measurements produced by these MRE methods are comparable. Hence, the purpose of this pilot study was to evaluate whether the measurements of the viscoelastic properties produced by 2D (stiffness) and 3D (elasticity and ‘Gabs,Elastic’) MRE are comparable.Materials and methodsSeven patients with diffuse or suspect diffuse liver disease were examined in the same day with the two MRE methods. 2D MRE was performed using an acoustic passive transducer, with a 1.5 T GE 450 W MR system. 3D MRE was performed using an electromagnetic active transducer, with a 1.5 T Philips Achieva MR system. Finally, mean viscoelastic values were extracted from the same anatomical region for both methods by an experienced radiologist.ResultsStiffness correlated well with the elasticity, R2 = 0.96 (P < 0.001; slope = 1.08, intercept = 0.61 kPa), as well as with ‘Gabs,Elastic’ R2 = 0.96 (P < 0.001; slope = 0.95, intercept = 0.28 kPa).ConclusionThis pilot study shows that different MRE methods can produce comparable measurements of the viscoelastic properties of the liver. The existence of such comparable measurements is important, both from a clinical as well as a research perspective, since it allows for equipment-independent monitoring of disease progression.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Liver; Rheology; Elastography; Fibrosis; MRE; MRI

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy