SwePub
Sök i LIBRIS databas

  Utökad sökning

onr:"swepub:oai:DiVA.org:liu-148844"
 

Sökning: onr:"swepub:oai:DiVA.org:liu-148844" > Evaluation of a hig...

Evaluation of a high framerate multi-exposure laser speckle contrast imaging setup

Hultman, Martin, 1992- (författare)
Linköpings universitet,Avdelningen för medicinsk teknik,Tekniska fakulteten
Fredriksson, Ingemar, 1980- (författare)
Linköpings universitet,Avdelningen för medicinsk teknik,Tekniska fakulteten,Perimed AB, Järfälla-Stockholm, Sweden
Strömberg, Tomas, 1966- (författare)
Linköpings universitet,Avdelningen för medicinsk teknik,Tekniska fakulteten
visa fler...
Larsson, Marcus, 1974- (författare)
Linköpings universitet,Avdelningen för medicinsk teknik,Tekniska fakulteten
visa färre...
 (creator_code:org_t)
SPIE - International Society for Optical Engineering, 2018
2018
Engelska.
Ingår i: High-Speed Biomedical Imaging and Spectroscopy III. - : SPIE - International Society for Optical Engineering. - 9781510614963
  • Konferensbidrag (refereegranskat)
Abstract Ämnesord
Stäng  
  • We present a first evaluation of a new multi-exposure laser speckle contrast imaging (MELSCI) system for assessing spatial variations in the microcirculatory perfusion. The MELSCI system is based on a 1000 frames per second 1-megapixel camera connected to a field programmable gate arrays (FPGA) capable of producing MELSCI data in realtime. The imaging system is evaluated against a single point laser Doppler flowmetry (LDF) system during occlusionrelease provocations of the arm in five subjects. Perfusion is calculated from MELSCI data using current state-of-the-art inverse models. The analysis displayed a good agreement between measured and modeled data, with an average error below 6%. This strongly indicates that the applied model is capable of accurately describing the MELSCI data and that the acquired data is of high quality. Comparing readings from the occlusion-release provocation showed that the MELSCI perfusion was significantly correlated (R=0.83) to the single point LDF perfusion, clearly outperforming perfusion estimations based on a single exposure time. We conclude that the MELSCI system provides blood flow images of enhanced quality, taking us one step closer to a system that accurately can monitor dynamic changes in skin perfusion over a large area in real-time

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
kon (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy