GEC-ESTRO ACROP recommendations on calibration and traceability of LE-LDR photon-emitting brachytherapy sources at the hospital level
-
- Perez-Calatayud, Jose (författare)
- Univ and Polytech La Fe Hosp, Spain; IIS La Fe UV, Spain
-
- Ballester, Facundo (författare)
- IIS La Fe UV, Spain; Univ Valencia, Spain
-
- Carlsson Tedgren, Åsa (författare)
- Karolinska Institutet,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Medicinsk strålningsfysik,Karolinska Univ Hosp, Sweden; Karolinska Inst, Sweden
-
visa fler...
-
- Rijnders, Alex (författare)
- Europe Hosp, Belgium
-
- Rivard, Mark J. (författare)
- Brown Univ, RI 02912 USA
-
- Andrassy, Michael (författare)
- RandD Brachytherapy Eckert and Ziegler BEBIG, Germany
-
- Niatsetski, Yury (författare)
- RandD Elekta Brachytherapy, Netherlands
-
- Schneider, Thorsten (författare)
- PTB, Germany
-
- Siebert, Frank-Andre (författare)
- UK S H, Germany
-
visa färre...
-
(creator_code:org_t)
- ELSEVIER IRELAND LTD, 2019
- 2019
- Engelska.
-
Ingår i: Radiotherapy and Oncology. - : ELSEVIER IRELAND LTD. - 0167-8140 .- 1879-0887. ; 135, s. 120-129
- Relaterad länk:
-
https://liu.diva-por... (primary) (Raw object)
-
visa fler...
-
http://www.thegreenj...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
http://kipublication...
-
visa färre...
Abstract
Ämnesord
Stäng
- Prostate brachytherapy treatment using permanent implantation of low-energy (LE) low-dose rate (LDR) sources is successfully and widely applied in Europe. In addition, seeds are used in other tumour sites, such as ophthalmic tumours, implanted temporarily. The calibration issues for LE-LDR photon emitting sources are specific and different from other sources used in brachytherapy. In this report, the BRAPHYQS (BRAchytherapy PHYsics Quality assurance System) working group of GEC-ESTRO, has developed the present recommendations to assure harmonized and high-quality seed calibration in European clinics. There are practical aspects for which a clarification/procedure is needed, including aspects not specifically accounted for in currently existing AAPM and ESTRO societal recommendations. The aim of this report has been to provide a European wide standard in LE-LDR source calibration at end-user level, in order to keep brachytherapy treatments with high safety and quality levels. The recommendations herein reflect the guidance to the ESTRO brachytherapy users and describe the procedures in a clinic or hospital to ensure the correct calibration of LE-LDR seeds. (C) 2019 The Authors. Published by Elsevier B.V.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- Brachytherapy; LE-LDR; Seeds; Calibration