Dose optimization for assessment of periodontal structures in cone beam CT examinations
-
- Al-Okshi, Ayman (författare)
- Malmö högskola,Odontologiska fakulteten (OD),Department of Oral Medicine and Radiology, Faculty of Dentistry, Sebha University, Sebha, Libya
-
- Theodorakou, Chrysoula (författare)
- Christie Medical Physics and Engineering, The Christie NHS Foundation Trust, Manchester Academic Health Science Centre, Manchester, UK
-
- Lindh, Christina (författare)
- Malmö högskola,Odontologiska fakulteten (OD)
-
(creator_code:org_t)
- British institute of radiology, 2016
- 2016
- Engelska.
-
Ingår i: Dento-Maxillo-Facial Radiology. - : British institute of radiology. - 0250-832X .- 1476-542X. ; 46:3
- Relaterad länk:
-
https://europepmc.or...
-
visa fler...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- OBJECTIVES: To investigate the relationship between dose and image quality for a dedicated dental CBCT scanner using different scanning protocols and to set up an optimal imaging protocol for assessment of periodontal structures. METHODS: Radiation dose and image quality measurements were made using 3D Accuitomo 170 (J. Morita, Kyoto, Japan) dental CBCT scanner. The SedentexCT IQ phantom was used to investigate the relationship between contrast-to-noise ratio (CNR) and dose-area product. Subjective image quality assessment was achieved using a small adult skull phantom for the same range of exposure settings. Five independent observers assessed the images for three anatomical landmarks using a three-point visual grade analysis. RESULTS: When correlating the CNR of each scanning protocol to the exposure parameters used to obtain it, CNR decreased as these parameters decreased, especially current-exposure time product. When correlating to subjective image quality, the CNR level remained acceptable when 5 mA and 17.5 s or greater was selected and 80 kV could be used without compromising the CNR. CONCLUSIONS: For a dedicated CBCT unit, changing the rotation angle from 360 degrees to 180 degrees degrades image quality. By altering tube potential and current for the 360 degrees rotation protocol, assessment of periodontal structures can be performed with a smaller dose without substantially affecting visualization.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Odontologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Dentistry (hsv//eng)