Sökning: onr:"swepub:oai:DiVA.org:miun-30340" >
Germanium "hexa" de...
Germanium "hexa" detector : Production and testing
-
- Sarajlić, M. (författare)
- Deutsches Elektronen-Synchrotron (DESY), Germany
-
- Pennicard, D. (författare)
- Deutsches Elektronen-Synchrotron (DESY), Germany
-
- Smoljanin, S. (författare)
- Deutsches Elektronen-Synchrotron (DESY), Germany
-
visa fler...
-
- Hirsemann, H. (författare)
- Deutsches Elektronen-Synchrotron (DESY), Germany
-
- Struth, B. (författare)
- Deutsches Elektronen-Synchrotron (DESY), Germany
-
- Fritzsch, T. (författare)
- Fraunhofer Institute for Reliability and Microintegration, Germany
-
- Rothermund, M. (författare)
- Fraunhofer Institute for Reliability and Microintegration, Germany
-
- Zuvic, M. (författare)
- Mirion Technologies (Canberra), France
-
- Lampert, M. O. (författare)
- Mirion Technologies (Canberra), France
-
- Askar, M. (författare)
- Al-Farabi Kazakh National University, Kazakhstan
-
- Graafsma, Heinz (författare)
- Mittuniversitetet,Avdelningen för elektronikkonstruktion,Deutsches Elektronen-Synchrotron (DESY), Germany
-
visa färre...
-
(creator_code:org_t)
- 2017
- 2017
- Engelska.
-
Ingår i: Journal of Instrumentation. - 1748-0221. ; 12:1
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Here we present new result on the testing of a Germanium sensor for X-ray radiation. The system is made of 3 × 2 Medipix3RX chips, bump-bonded to a monolithic sensor, and is called "hexa". Its dimensions are 45 × 30 mm2 and the sensor thickness was 1.5 mm. The total number of the pixels is 393216 in the matrix 768 × 512 with pixel pitch 55 μ m. Medipix3RX read-out chip provides photon counting read-out with single photon sensitivity. The sensor is cooled to -126°C and noise levels together with flat field response are measured. For -200 V polarization bias, leakage current was 4.4 mA (3.2 μ A/mm2). Due to higher leakage around 2.5% of all pixels stay non-responsive. More than 99% of all pixels are bump bonded correctly. In this paper we present the experimental set-up, threshold equalization procedure, image acquisition and the technique for bump bond quality estimate.
Ämnesord
- NATURVETENSKAP -- Fysik -- Acceleratorfysik och instrumentering (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Accelerator Physics and Instrumentation (hsv//eng)
Nyckelord
- Hybrid detectors
- Materials for solid-state detectors
- Solid state detectors
- X-ray detectors
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas
- Av författaren/redakt...
-
Sarajlić, M.
-
Pennicard, D.
-
Smoljanin, S.
-
Hirsemann, H.
-
Struth, B.
-
Fritzsch, T.
-
visa fler...
-
Rothermund, M.
-
Zuvic, M.
-
Lampert, M. O.
-
Askar, M.
-
Graafsma, Heinz
-
visa färre...
- Om ämnet
-
- NATURVETENSKAP
-
NATURVETENSKAP
-
och Fysik
-
och Acceleratorfysik ...
- Artiklar i publikationen
-
Journal of Instr ...
- Av lärosätet
-
Mittuniversitetet