Sökning: onr:"swepub:oai:DiVA.org:oru-111039" >
Assessing the perfo...
Assessing the performance of a targeted absolute quantification isotope dilution liquid chromatograhy tandem mass spectrometry assay versus a commercial nontargeted relative quantification assay for detection of three major perfluoroalkyls in human blood
-
- Salihovic, Samira, Associate Senior Lecturer, 1985- (författare)
- Örebro universitet,Institutionen för medicinska vetenskaper,Örebro Univ, Fac Med & Hlth, Sch Med Sci, Örebro, Sweden.
-
- Dunder, Linda (författare)
- Uppsala universitet,Arbets- och miljömedicin,Klinisk epidemiologi
-
- Lind, Monica, 1957- (författare)
- Uppsala universitet,Arbets- och miljömedicin
-
visa fler...
-
- Lind, Lars (författare)
- Uppsala universitet,Klinisk epidemiologi
-
visa färre...
-
(creator_code:org_t)
- John Wiley & Sons, 2024
- 2024
- Engelska.
-
Ingår i: Journal of Mass Spectrometry. - : John Wiley & Sons. - 1076-5174 .- 1096-9888. ; 59:2
- Relaterad länk:
-
https://doi.org/10.1...
-
visa fler...
-
https://uu.diva-port... (primary) (Raw object)
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
https://urn.kb.se/re...
-
visa färre...
Abstract
Ämnesord
Stäng
- Isotope dilution ultrahigh-performance liquid chromatography coupled to tandem mass spectrometry (UHPLC-MS/MS) is commonly used for trace analysis of polyfluoroalkyl and perfluoroalkyl substances (PFAS) in difficult matrices. Commercial nontargeted analysis of major PFAS where relative concentrations are obtained cost effectively is rapidly emerging and is claimed to provide comparable results to that of absolute quantification using matrix matched calibration and isotope dilution UHPLC-MS/MS. However, this remains to be demonstrated on a large scale. We aimed to assess the performance of a targeted absolute quantification isotope dilution LC-MS/MS assay versus a commercial nontargeted relative quantification assay for detection of three major PFAS in human blood. We evaluated a population-based cohort of 503 individuals. Correlations were assessed using Spearman's rank correlation coefficients (rho). Precision and bias were assessed using Bland-Altman plots. For perfluorooctane sulfonic acid, the median concentrations were 5.10 ng/mL (interquartile range [IQR] 3.50-7.24 ng/mL), the two assays correlated with rho 0.83. For perfluorooctanoic acid, the median concentrations were 2.14 ng/mL (IQR 1.60-3.0 ng/mL), the two assays correlated with rho 0.92. For perfluorohexanesulfonate, the median concentrations were 5.5 ng/mL (IQR 2.50-11.61 ng/mL), the two assays correlated with rho 0.96. The Bland-Altman statistical test showed agreement of the mean difference for the majority of samples (97-98%) between the two assays. Absolute plasma concentrations of PFAS obtained using matrix matched calibration and isotope dilution UHPLC-MS/MS show agreement with relative plasma concentrations from a nontargeted commercial platform by Metabolon. We observed striking consistency between the two assays when examining the associations of the three PFAS with cholesterol, offering additional confidence in the validity of utilizing the nontargeted approach for correlations with various health phenotypes.
Ämnesord
- NATURVETENSKAP -- Kemi -- Analytisk kemi (hsv//swe)
- NATURAL SCIENCES -- Chemical Sciences -- Analytical Chemistry (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Hälsovetenskap -- Arbetsmedicin och miljömedicin (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Health Sciences -- Occupational Health and Environmental Health (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Medicinsk bioteknologi -- Medicinsk bioteknologi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Medical Biotechnology -- Medical Biotechnology (hsv//eng)
Nyckelord
- PFAS
- cholesterol
- isotope dilution
- mass spectrometry
- metabolon
- nontargeted
- targeted
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas