SwePub
Sök i LIBRIS databas

  Utökad sökning

onr:"swepub:oai:DiVA.org:umu-135578"
 

Sökning: onr:"swepub:oai:DiVA.org:umu-135578" > Tibial component ro...

  • Gudnason, AsgeirUppsala universitet,Ortopedi (författare)

Tibial component rotation around the transverse axis measured by radiostereometry predicts aseptic loosening better than maximal total point motion A follow-up of 116 total knee arthroplasties after at least 15 years

  • Artikel/kapitelEngelska2017

Förlag, utgivningsår, omfång ...

  • 2017-03-07
  • Medical Journals Sweden AB,2017
  • electronicrdacarrier

Nummerbeteckningar

  • LIBRIS-ID:oai:DiVA.org:umu-135578
  • https://urn.kb.se/resolve?urn=urn:nbn:se:umu:diva-135578URI
  • https://doi.org/10.1080/17453674.2017.1297001DOI
  • https://urn.kb.se/resolve?urn=urn:nbn:se:uu:diva-323672URI

Kompletterande språkuppgifter

  • Språk:engelska
  • Sammanfattning på:engelska

Ingår i deldatabas

Klassifikation

  • Ämneskategori:ref swepub-contenttype
  • Ämneskategori:art swepub-publicationtype

Anmärkningar

  • Background and purpose - Maximal total point motion (MTPM) measured by radiostereometry (RSA) is widely used as a predictor of total knee arthroplasty (TKA) loosening. We compared the ability of different RSA measurements at different time points to predict loosening of tibial TKA components in the long term. Patients and methods - 116 TKAs in 116 patients were included in our analysis. 16 (14.8-17.4) years after surgery, 5 tibial components had been revised due to aseptic loosening. Receiver operating characteristic curves were calculated in order to investigate the specificity and sensitivity of different RSA parameters at different thresholds. Results - Rotation around the transverse (x-) axis measured 2 years postoperatively had the best predictive value of all parameters, with an area under the curve (AUC) of 80%. Using a threshold of 0.8 degrees, a specificity of 85% and a sensitivity of 50% were reached. The AUC for tibial component distal translation was 79% and it was 77% for proximal translation, whereas it was only 68% for MTPM. Interpretation - Rotation of the cemented tibial component around the transverse axis, proximal translation, and distal translation are slightly better at predicting aseptic loosening than MTPM, and tibial component migration measured after 2 years gives a good prediction of aseptic loosening up to 15 years.

Ämnesord och genrebeteckningar

Biuppslag (personer, institutioner, konferenser, titlar ...)

  • Adalberth, GunnarUppsala universitet,Ortopedi(Swepub:uu)gunnarab (författare)
  • Nilsson, Kjell-GunnarUmeå universitet,Institutionen för kirurgisk och perioperativ vetenskap,Univ Umea, Dept Surg & Perioperat Sci, Umea, Sweden.(Swepub:umu)kjni0001 (författare)
  • Hailer, Nils PUppsala universitet,Ortopedi(Swepub:uu)nilha837 (författare)
  • Uppsala universitetOrtopedi (creator_code:org_t)

Sammanhörande titlar

  • Ingår i:Acta Orthopaedica: Medical Journals Sweden AB88:3, s. 282-2871745-36741745-3682