Maintaining Image Quality While Reducing Acoustic Noise and Switched Gradient Field Exposure During Lumbar MRI

Glans, Anton (författare)
Umeå universitet,Institutionen för omvårdnad,Radiofysik
Wilén, Jonna (författare)
Umeå universitet,Radiofysik
Lindgren, Lenita (författare)
Umeå universitet,Institutionen för omvårdnad,Umeå centrum för funktionell hjärnavbildning (UFBI)
 (creator_code:org_t)
2021-02-09
2021
Engelska.
Ingår i: Journal of Magnetic Resonance Imaging. - : John Wiley & Sons. - 1053-1807 .- 1522-2586. ; 54:1, s. 315-325
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background: MR-generated acoustic noise can contribute to patient discomfort and potentially be harmful. One way to reduce this noise is by altering the gradient output and/or waveform using software optimization. Such modifications might influence image quality and switched gradient field exposure, and different techniques appear to affect sound pressure levels (SPLs) to various degrees.Purpose: To evaluate SPLs, image quality, switched gradient field exposure, and participants' perceived noise levels during two different acoustic noise reduction (ANR) techniques, Quiet Suite (QS) and Whisper Mode (WM), and to compare them with conventional T2-weighted turbo spin echo (T2W TSE) of the lumbar spine.Design: Prospective.Subjects: Forty adults referred for lumbar MRI.Field strength/sequence: Conventional T2W TSE, T2W TSE with QS, and T2W TSE with WM were acquired at 1.5 T.Assessment: Peak SPL (A-weighted decibels, dBA), perceived noise levels (Borg CR10®-scale), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), three radiologists' qualitative assessments in image quality on an ordinal scale 1-4, switched gradient field exposure (% general public), and gradient currents were measured. Interobserver reliability was reported as percentage agreement.Statistical tests: Repeated measures ANOVA, Friedman's ANOVA, and Wilcoxon's Signed-Rank Test for acoustic noise measurements and image quality assessments.Results: Mean peak SPLs were 89.9 dBA, 74.3 dBA, and 78.8 dBA for conventional, QS, and WM, respectively (P < 0.05). Participants perceived QS as the quietest and conventional as the loudest sequence (P < 0.05). No qualitative differences in image quality were seen (P > 0.05), although QS showed significantly improved SNR and CNR (P < 0.05). Switched gradient field exposure was reduced by 66% and 48% for QS and WM, respectively.Data conclusion: Without degrading image quality, both QS and WM are viable ANR techniques in lumbar T2W TSE. QS provided the lowest SPL, the lowest gradient field exposure and was perceived as the most silent among the three sequences.Level of evidence: 1 TECHNICAL EFFICACY STAGE: 5.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Hälsovetenskap -- Omvårdnad (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Health Sciences -- Nursing (hsv//eng)

Nyckelord

MRI safety
magnetic field
patient comfort
quiet suite
whisper mode

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Glans, Anton
Wilén, Jonna
Lindgren, Lenita
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Hälsovetenskap
och Omvårdnad
Artiklar i publikationen
Journal of Magne ...
Av lärosätet
Umeå universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy