SwePub
Sök i LIBRIS databas

  Utökad sökning

onr:"swepub:oai:DiVA.org:umu-39946"
 

Sökning: onr:"swepub:oai:DiVA.org:umu-39946" > Methods for the det...

Methods for the determination of effective monitor chamber thickness

Sjogren, R (författare)
Umeå universitet,Radiofysik
Karlsson, M G (författare)
Umeå universitet,Radiofysik
Karlsson, Mikael (författare)
Umeå universitet,Radiofysik
 (creator_code:org_t)
1999
1999
Engelska.
Ingår i: Medical physics (Lancaster). - 0094-2405. ; 26:9, s. 1871-1873
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • There are a number of models, both analytical and Monte Carlo, which are used to describe the fluence from the treatment head of accelerators. One common problem in these simulations is to find relevant information about details in the treatment head. A complex unit in the treatment head for which reliable data is seldom given is the monitor chamber. In this work two methods are described for obtaining this information by analyzing the increased scattering of an electron beam when the monitor chamber is introduced in the beam. It was found that the effective thickness of the electrodes in a monitor chamber can be determined with sufficient accuracy by using experimental results combined with Fermi-Eyges theory or Monte Carlo simulations. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)01009-3].

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Sjogren, R
Karlsson, M G
Karlsson, Mikael
Artiklar i publikationen
Medical physics ...
Av lärosätet
Umeå universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy