A Monte Carlo study of the dependence of early frame sampling on uncertainty and bias in pharmacokinetic parameters from dynamic PET
-
- Häggström, Ida, 1982- (författare)
- Umeå universitet,Radiofysik
-
- Axelsson, Jan (författare)
- Umeå universitet,Radiofysik
-
- Schmidtlein, Ross (författare)
- Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York 10065, USA
-
visa fler...
-
- Karlsson, Mikael (författare)
- Umeå universitet,Radiofysik
-
- Garpebring, Anders (författare)
- Umeå universitet,Radiofysik
-
- Johansson, Lennart (författare)
- Umeå universitet,Radiofysik
-
- Sörensen, Jens (författare)
- Uppsala universitet,Enheten för nuklearmedicin och PET
-
- Larsson, Anne (författare)
- Umeå universitet,Radiofysik
-
visa färre...
-
(creator_code:org_t)
- 2015-01-22
- 2015
- Engelska.
-
Ingår i: Journal of Nuclear Medicine Technology. - : Society of Nuclear Medicine. - 0091-4916 .- 1535-5675. ; 43:1, s. 53-60
- Relaterad länk:
-
http://tech.snmjourn...
-
visa fler...
-
https://urn.kb.se/re...
-
https://doi.org/10.2...
-
https://urn.kb.se/re...
-
visa färre...
Abstract
Ämnesord
Stäng
- Compartmental modeling of dynamic PET data enables quantifi- cation of tracer kinetics in vivo, through the calculated model parameters. In this study, we aimed to investigate the effect of early frame sampling and reconstruction method on pharmacokinetic parameters obtained from a 2-tissue model, in terms of bias and uncertainty (SD). Methods: The GATE Monte Carlo software was used to simulate 2 · 15 dynamic 3′-deoxy-3′-18F-fluorothymidine (18F-FLT) brain PET studies, typical in terms of noise level and kinetic parameters. The data were reconstructed by both 3- dimensional (3D) filtered backprojection with reprojection (3DRP) and 3D ordered-subset expectation maximization (OSEM) into 6 dynamic image sets with different early frame durations of 1, 2, 4, 6, 10, and 15 s. Bias and SD were evaluated for fitted parameter estimates, calculated from regions of interest. Results: The 2-tissue-model parameter estimates K1, k2, and fraction of arterial blood in tissue depended on early frame sampling, and a sampling of 6–15 s generally minimized bias and SD. The shortest sampling of 1 s yielded a 25% and 42% larger bias than the other schemes, for 3DRP and OSEM, respectively, and a parameter uncertainty that was 10%–70% higher. The schemes from 4 to 15 s were generally not significantly different in regards to bias and SD. Typically, the reconstruction method 3DRP yielded less framesampling dependence and less uncertain results, compared with OSEM, but was on average more biased. Conclusion: Of the 6 sampling schemes investigated in this study, an early frame duration of 6–15 s generally kept both bias and uncertainty to a minimum, for both 3DRP and OSEM reconstructions. Veryshort frames of 1 s should be avoided because they typically resulted in the largest parameter bias and uncertainty. Furthermore, 3DRP may be p
Ämnesord
- NATURVETENSKAP -- Fysik -- Annan fysik (hsv//swe)
- NATURAL SCIENCES -- Physical Sciences -- Other Physics Topics (hsv//eng)
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Image Processing (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- dynamic PET
- Monte Carlo; GATE
- compartment modeling
- frame sampling
- radiofysik
- radiation physics
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas