BACKGROUND:A computer controlled infusion pump (UIPump) for regulation of target tissue concentration of radioactive compounds was developed for use in biological research and tracer development for PET.METHODS:Based on observed tissue or plasma kinetics after a bolus injection of the tracer an algorithm calculates the infusion needed to obtain a specified target kinetic curve. A computer feeds this infusion scheme into an infusion pump connected to an animal via a venous catheter. The concept was validated using [11C]Flumazenil administrated to Sprague-Dawley rats where the whole brain distribution and kinetic of the tracer was measured over time using a microPET-scanner. The accuracy and precision of the system was assessed by producing steady-state levels of the tracer and by mimicking kinetics after oral administration.RESULTS:Various kinetic profiles could be generated, including rapid achievement of constant levels, or step-wise increased levels. The resulting tissue curves had low deviation from the target curves according to the specified criteria: AUC (%): 4.2 +/- 2.8, Maximal deviation (%): 13.6 +/- 5.0 and R2: 0.95 +/- 0.02.CONCLUSION:The UIPump-system is suitable for use in PET-research for assessment of PK/PD properties by simulation of different tracer tissue kinetics in vivo.
Ämnesord och genrebeteckningar
Biuppslag (personer, institutioner, konferenser, titlar ...)
-
Wallberg, Andreas
(författare)
-
Syvänen, StinaUppsala universitet,Institutionen för farmaceutisk biovetenskap,Hammarlund-Udenaes(Swepub:uu)stsyv838
(författare)
-
Josephsson, RaymondUppsala universitet,Klinisk virologi
(författare)
-
Långström, BengtUppsala universitet,Institutionen för biokemi och organisk kemi(Swepub:uu)benglang
(författare)
-
Bergström, MatsUppsala universitet,Institutionen för farmaceutisk biovetenskap
(författare)
-
Uppsala universitetEnheten för radiologi
(creator_code:org_t)
Sammanhörande titlar
-
Ingår i:BMC Medical Physics: Springer Science and Business Media LLC8:21756-6649
Internetlänk
Hitta via bibliotek
Till lärosätets databas