Performance evaluation of a particle-enhanced turbidimetric cystatin C assay on the Abbott ci8200 analyzer
-
- Flodin, Mats (författare)
- Uppsala universitet,Klinisk kemi,biokemisk struktur och funktion
-
- Larsson, Anders (författare)
- Uppsala universitet,Klinisk kemi,biokemisk struktur och funktion
-
(creator_code:org_t)
- Elsevier BV, 2009
- 2009
- Engelska.
-
Ingår i: Clinical Biochemistry. - : Elsevier BV. - 0009-9120 .- 1873-2933. ; 42:9, s. 873-876
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- OBJECTIVES: Glomerular filtration rate (GFR) is widely accepted as the best overall measure of kidney function. Cystatin C is a novel endogenous GFR marker that has been shown to be superior to creatinine for estimation of GFR in several studies. There is a need for cystatin C assays adapted to routine chemistry instrument to minimize turnaround times and allowing 24 h/day availability. MATERIALS AND METHODS: We have evaluated a new cystatin C assay developed for Architect cSystem (Abbott Laboratories, Abbott Park, IL, USA). RESULTS: The cystatin C assay showed good agreement with the corresponding assay from Dade Behring (Deerfield, IL, USA). The assay has a very low total imprecision and a good linearity. CONCLUSIONS: The new cystatin C assay is an interesting alternative to current cystatin C assays. On an Architect cSystem the assay can be performed with the same turnaround times and availability as creatinine.
Nyckelord
- MEDICINE
- MEDICIN
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas