Evaluation of Gentian cystatin C reagent on Abbott Ci8200 and calculation of glomerular filtration rate expressed in mL/min/1.73 m(2) from the cystatin C values in mg/L
-
- Flodin, Mats (författare)
- Uppsala universitet,Institutionen för medicinska vetenskaper,Biokemisk struktur och funktion
-
- Jonsson, A-S. (författare)
- Biokemisk struktur och funktion
-
- Hansson, L-O. (författare)
- Uppsala universitet,Institutionen för medicinska vetenskaper,Biokemisk struktur och funktion
-
visa fler...
-
- Danielsson, L-A. (författare)
- Biokemisk struktur och funktion
-
- Larsson, Anders (författare)
- Uppsala universitet,Klinisk kemi
-
visa färre...
-
(creator_code:org_t)
- 2009-07-08
- 2007
- Engelska.
-
Ingår i: Scandinavian Journal of Clinical and Laboratory Investigation. - : Informa UK Limited. - 0036-5513 .- 1502-7686. ; 67:5, s. 560-567
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- OBJECTIVE: Estimation of the glomerular filtration rate (GFR) is essential when evaluating patients with kidney disease and treating patients with drugs eliminated from the circulation by the kidneys. Cystatin C has been shown in several studies to be superior to creatinine in the estimation of GFR. At our hospitals, there is an increasing demand for cystatin C and at present we perform approximately 1500 cystatin C analyses a month. We thus need the assay available 24 h/day and to have it on our routine chemistry instrument to minimize handling time per test and time to reported test results. MATERIAL AND METHODS: We have evaluated a new cystatin C immunoassay from Gentian (Gentian, Moss, Norway) on Architect ci8200 (Abbott Laboratories, Abbott Park, Ill., USA). A prerequisite at our hospital is that cystatin C results are reported as a calculated GFR in mL/min/1.73 m(2), so we also made a comparison with iohexol clearance. RESULTS: The Gentian cystatin C assay showed good agreement with the corresponding assay from Dade Behring (Deerfield, Ill., USA) and good inter-laboratory concordance. The assay has very low total imprecision, good linearity and strong correlation with iohexol clearance (R (2) = 0.956). The equation for the correlation curve is: y = 79.901x(-1.4389). CONCLUSIONS: There was low inter-laboratory variation between the three laboratories involved in the cystatin C evaluation, and thus all three laboratories can use the same equation for calculating the estimated GFR.
Nyckelord
- Cystatin C
- glomerular filtration ratio
- human
- iohexol
- kidney
- MEDICINE
- MEDICIN